External ventricular drainage (EVD) is an emergency neurosurgical procedure to decrease intracranial pressure through a catheter mediated drainage of cerebrospinal fluid. Most EVD catheters are placed using free hands without direct visualization of the target and catheter trajectory, leading to a high rate of complications- hemorrhage, brain injury and suboptimal catheter placement. Use of stereotactic systems can prevent these complications. However, they have found limited application for this procedure due to their long set-up time and expensive hardware. Therefore, we have developed and pre-clinically validated a novel 3D printed stereotactic system for rapid and accurate implantation of EVD catheters. Its mechanical and imaging accuracies were found to be at par with clinical stereotactic systems. Preclinical trial in human cadaver specimens revealed improved targeting accuracy achieved within an acceptable time frame compared to the free hand technique. CT angiography emulated using cadaver specimen with radio-opaque vascular contrast showed vessel free catheter trajectory. This could potentially translate to reduced hemorrhage rate. Thus, our 3D printed stereotactic system offers the potential to improve the accuracy and safety of EVD catheter placement for patients without significantly increasing the procedure time.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8410845PMC
http://dx.doi.org/10.1038/s41598-021-97080-2DOI Listing

Publication Analysis

Top Keywords

stereotactic system
12
external ventricular
8
ventricular drainage
8
evd catheters
8
catheter trajectory
8
catheter placement
8
stereotactic systems
8
printed stereotactic
8
stereotactic
5
catheter
5

Similar Publications

Cost-minimization analysis of the GORTEC 2014-04 randomized phase II study of stereotactic ablative radiotherapy (SABR) or chemotherapy-SABR in oligometastatic head and neck cancer.

Radiother Oncol

January 2025

GORTEC 4 bis rue Emile Zola 37000 Tours, France; Department of Radiotherapy, Centre François-Baclesse, Corpuscular Physics Laboratory, IN2P3, ENSICAEN, CNRS UMR 6534, Université de Normandie, Caen, France. Electronic address:

Purpose: The randomized phase II GORTEC 2014-04 and French Head and Neck Intergroup study showed deeper deterioration of the quality of life (HRQoL) and dramatically higher severe toxicity rates with similar overall survival rates using chemo-SABR compared to SABR alone in oligometastatic head and neck cancer (HNSCC) patients. We evaluated the costs associated with SABR-alone versus chemo-SABR and their associated costs (transportation, hospitalizations, etc).

Materials And Methods: 69 HNSCC patients with 1-3 oligometastases and a controlled primary were randomized from September 2015 to October 2022.

View Article and Find Full Text PDF

Primary large B-cell lymphoma of the central nervous system misdiagnosed as autoimmune encephalitis: a case report.

Front Oncol

January 2025

Department of 2ndBrain Center and Stroke Center, The Affiliated Panyu Central Hospital, Guangzhou Medical University, Guangzhou, China.

Primary central nervous system lymphomas (PCNSL) are rare, constituting 2 - 3% of intracranial malignancies. A 49-year-old male presented with a 20-day history of dizziness and a 15-day history of right-sided weakness. Physical examination revealed various abnormal signs.

View Article and Find Full Text PDF

Potential for reduction of radiation dose in the assessment of the lead orientation in directional deep brain stimulation electrodes.

Eur J Radiol

January 2025

Department of Stereotactic and Functional Neurosurgery, Medical Center - University of Freiburg, Faculty of Medicine, 79106 Freiburg, Germany; Fraunhofer Institute for Laser Technology (ILT), 52074 Aachen, Germany.

Purpose: Directional deep brain stimulation (dDBS) relies on electrodes steering the stimulation field in a specific direction. Post implantation, however, the intended and real orientation of the lead frequently deviates e.g.

View Article and Find Full Text PDF

Purpose: Automatic planning (AP) has been compared to manual planning (MP) in lung stereotactic body radiation therapy (SBRT) to validate the former and to implement it in clinical practice.

Methods: A new developing Guided Planning System (GPS) engine was used to reoptimize 20 lung SBRT plans with the RayStation™ treatment planning system (TPS). The original manual plans were optimized to deliver 60 Gy in 5 or 8 fractions to the target with constraints on organs at risk (OARs) based on an internal protocol.

View Article and Find Full Text PDF

Background: Various methods exist to correct for intrafraction motion (IFM) of the prostate during radiotherapy. We sought to characterize setup corrections in our practice informed by the TrueBeam Advanced imaging package, and analyze factors associated with IFM.

Methods: 132 men received radiation therapy for prostate cancer with a volumetric modulated arc therapy technique.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!