Background: Patients with end-stage, biventricular heart failure, and for whom heart transplantation is not an option, may be given a Total Artificial Heart (TAH). The Realheart® is a novel TAH which pumps blood by mimicking the native heart with translation of an atrioventricular plane. The aim of this work was to create a strategy for using Computational Fluid Dynamics (CFD) to simulate haemodynamics in the Realheart®, including motion of the atrioventricular plane and valves.
Methods: The accuracies of four different computational methods for simulating fluid-structure interaction of the prosthetic valves were assessed by comparison of chamber pressures and flow rates with experimental measurements. The four strategies were: prescribed motion of valves opening and closing at the atrioventricular plane extrema; simulation of fluid-structure interaction of both valves; prescribed motion of the mitral valve with simulation of fluid-structure interaction of the aortic valve; motion of both valves prescribed from video analysis of experiments.
Results: The most accurate strategy (error in ventricular pressure of 6%, error in flow rate of 5%) used video-prescribed motion. With the Realheart operating at 80 bpm, the power consumption was 1.03 W, maximum shear stress was 15 Pa, and washout of the ventricle chamber after 4 cycles was 87%.
Conclusions: This study, the first CFD analysis of this novel TAH, demonstrates that good agreement between computational and experimental data can be achieved. This method will therefore enable future optimisation of the geometry and motion of the Realheart®.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/aor.14056 | DOI Listing |
J Cardiothorac Surg
December 2024
Department of Cardiology, Thomas Jefferson University, Philadelphia, PA, 19107, USA.
Background: Right ventricular (RV) function assessment by echocardiography can be challenging due to its complex morphology. Also, increasing use of sedation rather than general anesthesia for transfemoral approach transcatheter aortic valve replacement (TAVR) reduces the need for intraoperative transesophageal echocardiography (TEE). Recent clinical studies have demonstrated the importance of 3-dimensional (3D) echocardiography and a longitudinal strain for RV function assessment.
View Article and Find Full Text PDFEchocardiography
October 2024
Department of Ultrasonic Medicine, the First Affiliated Hospital of Guangxi Medical University, Nanning, China.
Bioengineering (Basel)
June 2024
California Medical Innovations Institute, San Diego, CA 92121, USA.
Adv Exp Med Biol
June 2024
Department of Medical Biology, Amsterdam Cardiovascular Sciences, Amsterdam University Medical Centres, Amsterdam, The Netherlands.
The formed hearts of vertebrates are widely different in anatomy and performance, yet their embryonic hearts are surprisingly similar. Developmental and molecular biology are making great advances in reconciling these differences by revealing an evolutionarily conserved building plan to the vertebrate heart. This suggests that perspectives from evolution may improve our understanding of the formation of the human heart.
View Article and Find Full Text PDFPediatr Cardiol
June 2024
Clinical Physiology, Department of Clinical Sciences Lund, Lund University, Lund, Sweden.
Hydraulic force aids diastolic filling of the left ventricle (LV) by facilitating basal movement of the atrioventricular plane. The short-axis atrioventricular area difference (AVAD) determines direction and magnitude of this force. Patients with atrial septal defect (ASD) have reduced LV filling due to the left-to-right shunt across the atrial septum and thus potentially altered hydraulic force.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!