Purpose: Intravoxel incoherent motion (IVIM) studies are performed with different acquisition protocols. Comparing them requires knowledge of echo time (TE) dependencies. The TE-dependence of the biexponential perfusion fraction f is well-documented, unlike that of its triexponential counterparts f and f and the biexponential and triexponential pseudodiffusion coefficients D , , and . The purpose was to investigate the TE-dependence of these parameters and to check whether the triexponential pseudodiffusion compartments are associated with arterial and venous blood.

Methods: Fifteen healthy volunteers (19-58 y; mean: 24.7 y) underwent diffusion-weighted imaging of the abdomen with 24 b-values (0.2-800 s/mm ) at TEs of 45, 60, 75, and 90 ms. Regions of interest (ROIs) were manually drawn in the liver. One set of bi- and triexponential IVIM parameters per volunteer and TE was determined. The TE-dependence was assessed with the Kruskal-Wallis test.

Results: TE-dependence was observed for f (P < .001), f (P = .001), and f (P < .001). Their median values at the four measured TEs were: f: 0.198/0.240/0.274/0.359, f : 0.113/0.139/0.146/0.205, f : 0.115/0.155/0.182/0.194. D, D , , and showed no significant TE-dependence. Their values were: diffusion coefficient D (10 mm /s): 9.45/9.63/9.75/9.41, biexponential D (10 mm /s): 5.26/5.52/6.13/5.82, triexponential (10 mm /s): 1.73/2.91/2.25/2.51, triexponential (mm /s): 0.478/1.385/0.616/0.846.

Conclusion: f and f show similar TE-dependence as f, ie, increase with rising TE; an effect that must be accounted for when comparing different studies. The diffusion and pseudodiffusion coefficients might be compared without TE correction. Because of the similar TE-dependence of f and f , the triexponential pseudodiffusion compartments are most probably not associated to venous and arterial blood.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.28996DOI Listing

Publication Analysis

Top Keywords

triexponential pseudodiffusion
12
echo time
8
triexponential
8
biexponential triexponential
8
intravoxel incoherent
8
incoherent motion
8
pseudodiffusion coefficients
8
pseudodiffusion compartments
8
compartments associated
8
001 001
8

Similar Publications

Purpose: To evaluate reproducibility and interlobar agreement of intravoxel incoherent motion (IVIM) quantification in the liver across field strengths and MR scanners with different gradient hardware.

Methods: Cramer-Rao lower bound optimization was performed to determine optimized monopolar and motion-robust 2D (b-value and first-order motion moment [M]) IVIM-DWI acquisitions. Eleven healthy volunteers underwent diffusion MRI of the liver, where each optimized acquisition was obtained five times across three MRI scanners.

View Article and Find Full Text PDF

Purpose: Intravoxel incoherent motion (IVIM) studies are performed with different acquisition protocols. Comparing them requires knowledge of echo time (TE) dependencies. The TE-dependence of the biexponential perfusion fraction f is well-documented, unlike that of its triexponential counterparts f and f and the biexponential and triexponential pseudodiffusion coefficients D , , and .

View Article and Find Full Text PDF

Background: Studies on intravoxel incoherent motion (IVIM) imaging are carried out with different acquisition protocols.

Purpose: To investigate the dependence of IVIM parameters on the B field strength when using a bi- or triexponential model.

Study Type: Prospective.

View Article and Find Full Text PDF

In the kidneys, there is both blood flow through the capillaries and flow of pre-urine through the tubuli and collecting ducts. We hypothesized that diffusion-weighted (DW) MRI measures both blood and pre-urine flow when using a tri-exponential intravoxel incoherent motion (IVIM) model. Our aim was to systematically investigate and optimize tri-exponential IVIM-analysis for the kidney and test its sensitivity to renal perfusion changes in humans.

View Article and Find Full Text PDF

Purpose: To investigate the reproducibility of diffusion-weighted imaging (DWI) with ultrahigh b-values, and analyze the age-related differences in normal prostates.

Materials And Methods: In all, 67 healthy participants were divided into three age groups (group A, 15-30 years; group B, 31-50 years; group C, ≥51 years), and underwent DWI scanning twice with 15 b-factors from 0 to 3000 at 3.0T.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!