Introduction: This study aimed to assess the bone cuts accuracy of a system for total knee arthroplasty including an active robotic arm. A second objective was to compare the accuracy among orthopaedic surgeons of different levels of experience.
Methods: Three orthopaedic surgeons cut 10 sawbone knees each. Planned and actual bone cuts were compared using computed tomography. Difference with respect to the planning was expressed as three position and three orientation errors following the anatomical planes. Statistical tests were performed to detect bias and compare surgeons.
Results: None of the 30 knees presented an outlier error, meaning an error ≥3 mm or ≥3°. The root-mean-square values of the 12 error types were below 0.8 mm or 0.8°, except for the femoral proximal-distal errors (1.7 mm) and the tibial anterior-posterior errors (1.4 mm). Biases were observed, particularly in femoral proximal-distal and tibial anterior-posterior positions. Median differences between surgeons were all lower than 0.8 mm and 0.5°, with statistically significant differences among surgeons in the femoral proximal-distal errors and the tibial anterior-posterior errors.
Conclusions: The system tested in this study achieved accurate bone cuts independently of the surgeon's level of experience. Biases were observed, suggesting that there might be options to improve the accuracy, particularly in proximal-distal position for the femur and in anterior-posterior position for the tibia.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8397104 | PMC |
http://dx.doi.org/10.3390/jcm10163714 | DOI Listing |
J Exp Orthop
January 2025
Department of Orthopaedic Surgery and Trauma University Center of Montpellier, University of Montpellier Montpellier France.
Purpose: Gap-balanced total knee arthroplasty (TKA) technique relies on initial ligament evaluation, particularly in patient-specific implantation using computer-assisted technologies. This cadaveric study aimed to compare the reproducibility and reliability of medial and lateral gap measurements between manual stress testing and dynamic ligament balancer.
Methods: Initial gap acquisitions were assessed from eight cadaveric knees (four specimens) during the same navigated TKA procedure by five differently skilled surgeons (three seniors and two juniors).
Meat Sci
December 2024
Scotland's Rural College, West Mains Road, UK.
Three-dimensional (3D) measurements extracted from beef carcass images were used to predict the weight of four saleable meat yield (SMY) traits (total SMY and the SMY of the forequarter, flank, and hindquarter) and four primal cuts (sirloin, ribeye, topside and rump). Data were collected at two UK abattoirs using time-of-flight cameras and manual bone out methods. Predictions were made for 484 carcasses, using multiple linear regression (MLR) or machine learning (ML) techniques.
View Article and Find Full Text PDFClin Adv Periodontics
January 2025
Department of Orthodontics and Dentofacial Orthopedics, Eastman Institute for Oral Health, University of Rochester, Rochester, New York, USA.
Background: Gingival recession defects (GRDs) pose functional and esthetic concerns and may be associated with unfavorable tooth positions. Surgically facilitated orthodontic treatment (SFOT) with clear aligners can be a valuable option for adults with severe malocclusion and GRDs.
Methods: A 28-year-old male presented with severe dental crowding, Class III dental malocclusion, localized tooth crossbites, and tapered maxillary arch.
JBJS Essent Surg Tech
December 2024
Department of Orthopaedic Surgery, Cleveland Clinic Foundation, Cleveland, Ohio.
Background: Unicompartmental knee arthroplasty (UKA) procedures have become much more common in the United States in recent years, with >40,000 UKAs performed annually. However, it is estimated that 10% to 40% of UKAs fail and thus require conversion to total knee arthroplasty (TKA). In the field of total joint arthroplasty, robotic-assisted surgeries have demonstrated advantages such as better accuracy and precision of implant positioning and improved restoration of a neutral mechanical axis.
View Article and Find Full Text PDFBackground: Knee arthroplasty procedures improve pain, function, stability, and appearance of the limb. Total knee arthroplasty (TKA) in severe, long-standing osteoarthritis (OA) with large medial tibial defects could be a challenge. This paper looks at TKA outcomes when large tibial defects are managed without metal wedges or stems.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!