Purpose: To assess refractive and visual outcomes following phacoemulsification with femtosecond laser-assisted cataract surgery (FLACS) using intraoperative aberrometry and implantation of a toric intraocular lens (IOL) in eyes with different degrees of astigmatism.
Methods: One hundred two eyes of 70 patients who underwent implantation of the transitional toric 565 Precizon IOL (Ophtec BV) were enrolled. FLACS, capsular tension ring insertion, and intraoperative aberrometry were performed. Main outcome measures were refractive error, uncorrected- and corrected distance snellen decimal visual acuity values (UDVA and CDVA, respectively), and IOL rotation. Specifically, a vector analysis was carried out with J0 and J45 evaluation. Eyes were evaluated 1-year after surgery.
Results: Overall, 94.12% (96 eyes) and 100% (102 eyes) of the eyes showed a spherical equivalent (SE) within ± 0.50D and ± 1.00D, respectively. The mean SE and refractive cylinder were - 0.06 ± 0.29D and - 0.23 ± 0.37D, respectively. Vector analysis revealed that 100% of the eyes were within ± 0.50D for the J0 and J45 cylindrical components. The mean toric axis rotation was 1.10 ± 1.71° (from 0° to 5°), 77% (79 eyes), and 100% (102 eyes) of the eyes showed UDVA and CDVA of 20/25, respectively. The postoperative mean values of monocular UDVA and CDVA were 0.88 ± 0.17 and 0.96 ± 0.07 (about 20/20), respectively. No patient required IOL realignment during the postoperative follow-up.
Conclusions: The present study suggests that the use of the Precizon IOL after FLACS, using intraoperative aberrometry in patients with different amounts of astigmatism, provides good visual acuity, accurate refractive outcomes, and excellent rotational stability.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s10792-021-02012-8 | DOI Listing |
Indian J Ophthalmol
January 2025
Cornea, Cataract and Refractive Surgery Services, Dr. Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences, New Delhi, India.
Indian J Ophthalmol
January 2025
Dr. Rajendra Prasad Centre for Ophthalmic Sciences, AIIMS, New Delhi, India.
Scleral-fixated intraocular lens (SFIOL) is a widely used technique for IOL implantation in patients where capsular support is insufficient. Most surgeons have shifted away from sutured to sutureless SFIOL techniques where haptics of a multifocal IOL are inserted in scleral tunnels/flaps. Large-scale publications have shown wide variation in the refractive status of eyes post-SFIOL even in the best of the hands.
View Article and Find Full Text PDFClin Ophthalmol
December 2024
Alcon Vision LLC, Fort Worth, TX, USA.
Purpose: To compare the refractive prediction accuracy of the Optiwave Refractive Analysis (ORA) SYSTEM with the Barrett True-K (BTK) formula in calculating intraocular lens (IOL) power in eyes that underwent cataract surgery after previous myopic photorefractive keratectomy (PRK) or laser-assisted in situ keratomileusis (LASIK).
Methods: This retrospective study evaluated patients aged ≥22 years with prior myopic PRK or LASIK who underwent unilateral or bilateral cataract removal and monofocal IOL implantation using the ORA SYSTEM at 177 sites in the United States. Two datasets were analyzed: All Eyes (ie, all eligible eyes) and First Surgery Eyes (ie, each patient's first implanted eye).
Clin Ophthalmol
November 2024
PersonalEyes, Sydney, NSW, Australia.
Purpose: To assess ocular higher-order aberrations (HOAs) following ray trace guided laser in situ keratomileusis (LASIK).
Methods: Retrospective review at a single-site ophthalmology practice. Two hundred and fifty eyes of 250 patients with myopia and astigmatism undergoing ray trace LASIK were randomly reviewed (Wavelight Plus, Alcon).
Front Med (Lausanne)
October 2024
Optics and Optometry and Vision Sciences Department, University of Valencia, Valencia, Spain.
Background: To compare the accuracy of intraoperative wavefront aberrometry using the ORA VLynk system with different biometry-based formulas in short and long eyes after cataract surgery.
Methods: This prospective study considered 48 eyes with axial lengths of <22.1 mm and 48 eyes with axial lengths of >25.