Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: An N-terminal octapeptide cleavage of the cystatin C protein was discovered by mass spectrometry when cerebrospinal fluid (CSF) was stored at -20°C for 3 months, which did not occur when CSF was stored at -80°C.
Objective: The aim was to develop an immunoassay as quality assessment tool to detect this -20°C cleavage of cystatin C in CSF and support Alzheimer's disease research.
Methods: A specific monoclonal antibody and a double indirect sandwich ELISA were developed: one assay quantifies the octapeptide uncleaved protein specifically and the other quantifies the total cystatin C present in the biological fluid (both cleaved and uncleaved forms). The ratio of these concentrations was calculated to assess the extent of cleavage of cystatin C. The novel ELISA was validated and applied in a short-term (up to 4 weeks) and mid-term (up to one year) stability study of CSF stored at 4°C, -20°C, -80°C, and liquid nitrogen. Impact of freeze-thaw cycles, adsorption, and protease inhibitors were tested.
Results: The ratio of truncated protein was modified following -20°C storage and seemed to reach a plateau after 6 months. The ratio was impacted neither by freeze-thaw cycles nor adsorption. The -20°C specific cleavage was found to be protease related.
Conclusion: Using this novel double indirect sandwich ELISA, absolute levels of the total and uncleaved cystatin C and the ratio of truncated cystatin C can be measured. This assay is an easily applicable tool which can be used to confirm that CSF biospecimen are fit-for-purpose for Alzheimer's disease research.