Purpose: Photo-optical TCpO2 (pTCpO2) has been proposed as a new method to determine the partial oxygen pressure of the lower extremity in patients with peripheral arterial disease. It is aimed to determine the level of agreement between pTCpO2 and the traditional electro-chemical transcutaneous oxygen tension measurement (eTCpO2).

Methods: Eighteen patients with intermittent claudication underwent simultaneous ankle-brachial index measurement, toe-pressure, pTCpO2 and eTCpO2 tests. Oxygen tension levels were measured on anterior chest and calf prior in rest (T0), during induced ischemia (T1) and after blood flow restoration (T2). TCpO2 agreement was assessed according to the principles of Bland and Altman.

Results: Absolute average TCpO2 values differed between eTCpO2 and pTCpO2 for calf in T2 (38,1 mmHg (σ 14,4) vs. 49,8 (σ 22.3) with P = 0.35). The Bland-Altman plots demonstrated eTCpO2 and pTCpO2 bias of 3,7 mmHg (σ 18,8), 11,6 mmHg (σ 17,6) and 6,7 mmHg (σ 23,5) for T0, T1 and T2 for the calf.

Conclusion: pTCpO2 is in agreement with eTCpO2 in measuring pO2 levels of the lower extremity in rest and during induced ischemia in patients with vascular claudication. The large variability between eTCpO2 and pTCpO2 should be accounted for, while pTCpO2 values have a tendency to demonstrate higher values in comparison to eTCpO2.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.avsg.2021.05.019DOI Listing

Publication Analysis

Top Keywords

oxygen tension
16
transcutaneous oxygen
12
tension measurement
12
etcpo2 ptcpo2
12
electro-chemical transcutaneous
8
ptcpo2
8
lower extremity
8
rest induced
8
induced ischemia
8
etcpo2
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!