0° vs. 180° CT localiser: The effect of vertical off-centring, phantom positioning and tube voltage on dose optimisation in multidetector computed tomography.

J Med Radiat Sci

Medical Radiation Science, School of Dentistry and Health Sciences, Faculty of Science and Health, Charles Sturt University, Wagga Wagga, New South Wales, Australia.

Published: March 2022

Introduction: Patient positioning is an essential consideration for the optimisation of radiation dose during CT examinations. The study objectives seek to explore the effects of vertical off-centring, localiser direction (0° and 180°), and phantom positioning (supine and prone) on radiation dose, using three different tube voltages in multidetector computed tomography (MDCT) imaging.

Methods: The trunk of a PBU-60 anthropomorphic phantom was imaged using a Discovery CT750 HD - 128 slice (GE Healthcare). Images employing 0° and 180° localisers were acquired in supine and prone orientation for each combination of vertical off-centring (±100, ±60 and ±30 mm) and different tube voltages (80, 120 and 140 kVp), using the system's automatic tube current modulation (ATCM) function. The displayed volume CT dose index (CTDI ) and dose length product (DLP) were recorded.

Results: With incremental table off-centring of ±100 mm, the dose at 120 kVp in the supine position ranged from 63% to 196% (0° localiser) and from 66% to 191% (180° localiser) as compared to iso-centre. While in the prone position, the dose ranged from 62% to 195% (0° localiser); and 62% to 193% (180° localiser), with a notable dose increase at higher tube voltages. Dose variation and vertical off-centring showed a significant relationship for both 0° and 180° localisers (r = 0.94 and 0.96, respectively, P < 0.001). The CTDI variation between supine and prone phantom positions at ±100 mm off-centring was 0.22 mGy (2.9%), and 0.19 mGy (2.3%) when the 0° and 180 ° localisers were utilised, respectively.

Conclusions: Phantom off-centring and localiser direction evidenced large dose variation. It is recommended that the 0° localiser is employed during CT examinations, in order to minimise the potential additional radiation dose which may result from off-centring and the use of lower tube voltages where clinically appropriate.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8892417PMC
http://dx.doi.org/10.1002/jmrs.535DOI Listing

Publication Analysis

Top Keywords

0° 180°
16
vertical off-centring
16
180° localiser
12
tube voltages
12
dose
9
phantom positioning
8
multidetector computed
8
computed tomography
8
radiation dose
8
supine prone
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!