Fracture of a 40-mm BIOLOX Delta Femoral Head.

Arthroplast Today

Department of Orthopaedics, Duke University, Durham, NC, USA.

Published: August 2021

Ceramic bearing surfaces have gained popularity in total hip arthroplasty as a result of the favorable mechanical properties and low wear rates. Despite the recognition as an attractive articulation, problems such as ceramic head fracture persist. Smaller heads and higher body mass indices are touted as risk factors for ceramic head fracture and are often associated with antecedent trauma. We present a case report of an 83-year-old male with a body mass index of 26.7 kg/m who suffered a fracture of a 40-mm ceramic femoral head. This occurred atraumatically 5 years from his index surgery. This patient underwent revision total hip arthroplasty which included debridement of ceramic debris and alteration of the bearing surface with femoral head and polyethylene liner exchange.