Impact of deformation on a supine-positioned image-guided breast surgery approach.

Int J Comput Assist Radiol Surg

Department of Biomedical Engineering, Vanderbilt University, 1225 Stevenson Center Ln, Nashville, 37235, USA.

Published: November 2021

Purpose: To reduce reoperation rates for image-guided breast-conserving surgery, the enhanced sensitivity of magnetic resonance (MR) supine imaging may be leveraged. However, accurate tissue correspondence between images and their physical counterpart in the surgical presentation is challenging due to breast deformations (e.g., from patient/arm position changes, and operating room table rotation differences). In this study, standard rigid registration methods are employed and tissue deformation is characterized.

Methods: On n = 10 healthy breasts, surface displacements were measured by comparing intraoperative fiducial locations as the arm was moved from conventional MR scanning positions (arm-down and arm-up) to the laterally extended surgical configuration. Supine MR images in the arm-down and arm-up positions were registered to mock intraoperative presentations.

Results: Breast displacements from a supine MR imaging configuration to a mock surgical presentation were 28.9 ± 9.2 mm with shifts occurring primarily in the inferior/superior direction. With respect to supine MR to surgical alignment, the average fiducial, target, and maximum target registration errors were 9.0 ± 1.7 mm, 9.3 ± 1.7 mm, and 20.0 ± 7.6 mm, respectively. Even when maintaining similar arm positions in the MR image and mock surgery, the respective averages were 6.0 ± 1.0 mm, 6.5 ± 1.1 mm, and 12.5 ± 2.8 mm.

Conclusion: From supine MR positioning to surgical presentation, the breast undergoes large displacements (9.9-70.1 mm). The data also suggest that significant nonrigid deformations (9.3 ± 1.7 mm with 20.0 mm average maximum) exist that need to be considered in image guidance and modeling applications.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9237955PMC
http://dx.doi.org/10.1007/s11548-021-02452-8DOI Listing

Publication Analysis

Top Keywords

surgical presentation
12
supine imaging
8
arm-down arm-up
8
supine
5
surgical
5
impact deformation
4
deformation supine-positioned
4
supine-positioned image-guided
4
breast
4
image-guided breast
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!