Objective: The long-term durability of transcatheter aortic bioprosthetic valves continues to be a major concern. Standardized criteria of the structural valve deterioration (SVD) and bioprosthetic valve failure (BVF) have recently been defined. Limited studies have evaluated the long-term durability of transcatheter aortic valve implantation (TAVI) according to these new definitions. We aim to analyze the durability of TAVI beyond 5 years and to report the frequency of SVD and BVF.
Methods: A total of 89 patients who had undergone TAVI and had theoretically completed at least 5 years after the procedure were included. Either a Medtronic CoreValve or an Edwards SAPIEN XT valve were implanted in the patients. New standardized definitions were used to evaluate SVD and BVF.
Results: The mean age of the patients was 78.70±6.95 years. SVD occured in 4 (4.5%) patients during 6 years of follow-up. Severe SVD was observed in 2 patients (2.2%), and these patients had the New York Heart Association class II symptoms. Both patients with severe SVD also met the criteria of BVF. Moderate SVD was observed in 2 patients (2.2%), and these patients had no valve-related symptoms. Of the 4 SVD cases, 2 were associated with increased mean transaortic gradients, whereas the remaining 2 cases were associated with intraprosthetic aortic regurgitation. All patients with SVD are still alive, and none of them have required aortic valve reintervention.
Conclusion: Although first-generation TAVI devices were used, we determined the low rate of SVD and BVF at the 6-year follow-up. It may be suggested that there is no major concern associated with TAVI even with first-generation devices regarding long-term durability.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8357435 | PMC |
http://dx.doi.org/10.5152/AnatolJCardiol.2021.37670 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!