Objectives: Trimethylaminuria, also known as Fish Odor Syndrome (FOS), is a condition characterized by the presence of high concentrations of trimethylamine (TMA) in urine, sweat and expired air of affected patients. Diagnosis of this benign but unpleasant disease is mainly based on clinical presentation and assessment of TMA and its metabolite, TMAO (trimethylamine-N-oxide), concentrations in urine of patients.

Material And Methods: We here described the validation of an analytical method for measurement of TMA and TMAO in urine using nuclear magnetic resonance (NMR) according to the specifications of the ISO 15189 norm. We used a fast validation protocol, based exactitude profile method, enabling to determine accuracy, intra and inter-day precision from a limited number of samples.

Results: The linearity was established from 2.5 to 100 mg/L for TMA measurement and from 10 to 1000 mg/L for TMAO measurement, with good analytical performances i.e. accuracy, intra and inter-day precision. We also report a case diagnose for FOS from this method.

Conclusions: This method validation ensures the robustness of NMR in routine use for diagnosis of trimethylaminuria, as part of the reference center for inherited metabolic diseases at the Tours hospital.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ab.2021.114330DOI Listing

Publication Analysis

Top Keywords

tma tmao
8
tmao measurement
8
nuclear magnetic
8
magnetic resonance
8
fish odor
8
odor syndrome
8
accuracy intra
8
intra inter-day
8
inter-day precision
8
tma
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!