Purpose: Feasability of a no-reference image quality metric was assessed on patient-like images using a patient-specific phantom simulating a frame of a coronary angiogram.
Methods: One background and one contrast-filled frame of a coronary angiogram, acquired using a clinical imaging protocol, were selected from a Philips Integris Allura FD (Philips Healthcare, Best, The Netherlands). The background frame's pixels were extruded to a thickness proportional to their grey value. One phantom was 3D printed using composite 80% bronze filament (max. thickness of 5.1 mm), the other was a custom PMMA cast (max thickness of 8.5 cm). A vessel mold was created from the contrast-filled frame and injected with a solution of 320 mg I/ml contrast fluid (75%), water and gelatin. Still X-ray frames of the vessel mold + background phantom + 16 cm PMMA were acquired at manually selected different exposure settings using a Philips Azurion (Philips Healthcare, Best, The Netherlands) in User Quality Control Mode and were exported as RAW images. The signal-difference-to-noise-ratio-squared (SDNR) and a spatial-domain-equivalent of the noise equivalent quanta (NEQ) were calculated. The Spearman's correlation of the latter parameters with a no-reference perceptual image quality metric (NIQE) was investigated.
Results: The bronze phantom showed better resemblance to the original patient frame selected from a coronary angiogram of an actual patient, with better contrast and less blur than the PMMA phantom. Both phantoms were imaged using a comparable imaging protocol to the one used to acquire the original frame. The bronze phantom was hence used together with the vessel mold for image quality measurements on the 165 still phantom frames. A strong correlation was noted between NEQ and NIQE (SROCC = -0.99, p < 0.0005) and between SDNR and NIQE (SROCC = -0.97, p < 0.0005).
Conclusion: Using a cost-effective and easy to realize patient-specific phantom we were able to generate patient-like X-ray frames. NIQE as a no-reference image quality model has the potential to predict physical image quality from patient images.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ejmp.2021.07.011 | DOI Listing |
Clin Exp Optom
January 2025
Department of Optometry, School of Paramedical and Rehabilitation Sciences, Mashhad University of Medical Sciences, Mashhad, Iran.
Clinical Relevance: The vestibular-ocular reflex stabilises the retinal image and maintains balance during head movement. Astigmatism is one of the common refractive errors that can reduce the quality of visual inputs.
Background: The purpose of this study was to investigate the effect of induced astigmatism on the function of the vestibular-ocular reflex.
Extracell Vesicles Circ Nucl Acids
December 2024
Department of Otorhinolaryngology, Head and Neck Surgery, The Chinese University of Hong Kong, Hong Kong Special Administrative Region, Hong Kong 00000, China.
Current approaches to oral cancer diagnosis primarily involve physical examination, tissue biopsy, and advanced computer-aided imaging techniques. However, despite these advances, patient survival rates have not significantly improved. Hence, there is a critical need to develop minimally invasive tools with high sensitivity and specificity to improve patient survival and quality of life.
View Article and Find Full Text PDFMethodsX
June 2025
Department of Computer Science and Engineering, Vel Tech Rangarajan Dr.Sagunthala R&D Institute of Science and Technology, Tamil Nadu 600062, India.
The disease affects the optic nerve and represents the principle reasons of irreversible vision loss, mostly asymptomatic and uncontrolled. Consequently, early and accurate diagnosis is critical to prevent or reduce its effect, however, conventional diagnostic techniques often fail to provide concrete results. In this regard, we present a new approach built on Generative Adversarial Networks (GAN) and MobileNetV2 pretrained architecture for diagnosing glaucoma.
View Article and Find Full Text PDFJPRAS Open
March 2025
Department of Radiation Oncology, Tri-Service General Hospital, National Defense Medical Center, Taipei, Taiwan.
Background: Breast cancer patients experience acute radiation dermatitis (ARD) during radiation therapy (RT). This study investigated the prophylactic effect of a newly developed xenogeneic platelet-rich plasma (PRP) lotion on ARD for breast cancer patients.
Methods: This study enrolled patients with ductal carcinoma in situ and early-stage invasive breast cancers after breast-conserving surgery.
Eur J Radiol Open
June 2025
Department of Radiology, The First Affiliated Hospital of Anhui Medical University, No.218 Jixi Road, Hefei, Anhui 230022, China.
Objectives: Subclinical myocardial involvement is common in systemic lupus erythematosus (SLE), but differences between new onset and longstanding SLE are not fully elucidated. This study compared myocardial involvement in new onset versus longstanding SLE using cardiovascular magnetic resonance (CMR).
Materials And Methods: We prospectively enrolled 24 drug-naïve new onset SLE patients, 27 longstanding SLE patients, and 20 healthy controls.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!