Occupational radiation exposure to the lens of the eye in interventional radiology.

J Med Imaging Radiat Oncol

Department of Radiology, Alfred Hospital, Melbourne, Victoria, Australia.

Published: February 2022

Introduction: Cataract formation is a tissue reaction effected by radiation exposure. The purpose of this study was to evaluate the occupational exposure to the lens of the eye of interventional radiologists (IR's) and interventional radiology staff, with and without lead glasses.

Methods: Ethical approval was provided by the hospital research and ethics committee. A prospective cohort study was performed over 1 year, doses recorded, lifetime dose (estimated at working 5 days in angiography, for 30 years) was estimated and dose compared to current guidelines. Thermoluminescent dosimeters (TLDs; Landauer, Glenwood, USA) Hp(3) were placed on both the exterior and interior side of the personal lead glasses worn by three interventional radiologists and two radiographers. They were monitored during all procedures performed within 1 year. Lead glasses (AttenuTech Microlite , Florida, USA) with specifications were 0.75 mm lead equivalent front shield, and Side shield 0.3 mm Pb equivalent. A control TLD was placed in the storage location of the lead glasses when not in use. Yearly dose was measured and lifetime dose was calculated from the data obtained. Calculation of dose received per day(s) spent performing procedures for both annual and lifetime exposure was performed. In addition a record of occurrence of splashes on glasses was made after each case.

Results: Eye doses without protection were double the recommended limits for both annual and lifetime dose. For interventional radiologists working between 3 and 4 or more days in the lab per week, annual dose thresholds would be exceeded (20 mSv/year averaged over 5 years, no more than 50 mSv in 1 year). If interventional radiologists worked between 3 and 4 or more days in the lab, lifetime dose thresholds would be exceeded (500 mSv lifetime dose). Lead glasses reduced radiation exposure by an average of 79%. If lead glasses were worn no interventional radiologists would exceed annual or lifetime dose thresholds to the eyes even if working 5 days per week as the primary operator. Radiographers would not exceed annual or lifetime dose thresholds even without lead glasses. Splash incidents occurred for all interventional radiologists and one radiographer.

Conclusion: The use of lead glasses even in this small study resulted in a decreased dose of radiation to the lens of the eye. Regular use of radiation protection eyewear will reduce eye dose for primary proceduralists to well below yearly and lifetime thresholds.

Download full-text PDF

Source
http://dx.doi.org/10.1111/1754-9485.13307DOI Listing

Publication Analysis

Top Keywords

lifetime dose
28
lead glasses
28
interventional radiologists
24
annual lifetime
16
dose thresholds
16
dose
13
radiation exposure
12
lens eye
12
lead
9
lifetime
9

Similar Publications

Purpose: Oral corticosteroids (OCS) are recommended for the treatment of exacerbations in people with COPD; however, high cumulative lifetime doses (≥1000mg prednisolone-equivalent) are associated with adverse health effects. This issue is well defined in asthma but is less well understood in COPD. The aim of this study was to examine cumulative OCS dispensed to people with COPD over 12 months.

View Article and Find Full Text PDF

Background: Total-body (TB) Positron Emission Tomography (PET) is one of the most promising medical diagnostics modalities, opening new perspectives for personalized medicine, low-dose imaging, multi-organ dynamic imaging or kinetic modeling. The high sensitivity provided by total-body technology can be advantageous for novel tomography methods like positronium imaging, demanding the registration of triple coincidences. Currently, state-of-the-art PET scanners use inorganic scintillators.

View Article and Find Full Text PDF

A health economic evaluation of the multinational, randomized controlled CONVINCE trial - cost-utility of high-dose online hemodiafiltration compared to high-flux hemodialysis.

Kidney Int

January 2025

Department of Epidemiology and Health Economics, Julius Center for Health Sciences and Primary Care, University Medical Center Utrecht, Utrecht University, Utrecht, The Netherlands.

High-flux hemodialysis (HD) and high-dose hemodiafiltration (HDF) are established treatments for patients with kidney failure. Since HDF has been associated with improved survival rates compared to HD, we evaluated the cost-effectiveness of HDF compared to HD. Cost-utility analyses were performed from a societal perspective alongside the multinational randomized controlled CONVINCE trial.

View Article and Find Full Text PDF

This study assesses the activity concentrations of the radionuclides 238U, 232Th, and 40K in soil samples collected from Wolaita Sodo town, located in the Southern Nations, Nationalities, and Peoples' (SNNP) Region, Ethiopia. A gamma-ray spectrometer equipped with a NaI(Tl) detector was used for the measurements. The concentrations of 238U, 232Th, and 40K varied from 3.

View Article and Find Full Text PDF

Whether early life acetaminophen (APAP) exposures injure the developing lung is controversial. We sought to correlate murine pulmonary developmental expression profiles of to susceptibility to APAP exposure. P14 C57BL/6 mice were exposed to APAP (140 mg/kg x 1, IP) and assessed for evidence of a histologic, metabolic, functional, and/or transcriptional pulmonary response.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!