Purpose: Aortic dissection (AD) is a catastrophic disease with complex hemodynamic conditions, however, understandings regarding its perfusion characteristics were not sufficient. In this study, a mock circulation loop (MCL) that integrated the Windkessel element and patient-specific silicone aortic phantoms was proposed to reproduce the aortic flow environment in vitro.
Materials And Methods: Patient-specific normal and dissected aortic phantoms with 12 branching vessels were established and embedded into this MCL. Velocities for aortic branches based on 20 healthy volunteers were regarded as the standardized data for flow division. By altering boundary conditions, the proposed MCL could mimic normal resting and left-sided heart failure (LHF) conditions. Flow rates and pressure status of the aortic branches could be quantified by separate sensors.
Results: In normal resting condition, the simulated heart rate and systemic flow rate were 60 bpm and 4.85 L/minute, respectively. For the LHF condition, the systolic and diastolic blood pressures were 75.94±0.77 mmHg and 57.65±0.35 mmHg, respectively. By tuning the vascular compliance and peripheral resistance, the flow distribution ratio (FDR) of each aortic branch was validated by the standardized data in the normal aortic phantom (mean difference 2.4%±1.70%). By comparing between the normal and dissected aortic models under resting condition, our results indicated that the AD model presented higher systolic (117.82±0.60 vs 108.75±2.26 mmHg) and diastolic (72.38±0.58 vs 70.46±2.33 mmHg) pressures, the time-average velocity in the true lumen (TL; 36.95 cm/s) was higher than that in the false lumen (FL; 22.95 cm/s), and the blood transport direction between the TL and FL varied in different re-entries.
Conclusions: The proposed MCL could be applied as a research tool for in vitro hemodynamic analysis of the aorta diseases under various physical conditions.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1177/15266028211034863 | DOI Listing |
Port J Card Thorac Vasc Surg
January 2025
Department of Biomedicine - Unit of Anatomy, Faculty of Medicine, University of Porto; RISE@Health, Porto, Portugal.
Background: Aortoiliac disease (AID) is a variant of peripheral artery disease involving the infrarenal aorta and iliac arteries. Similar to other arterial diseases, aortoiliac disease obstructs blood flow through narrowed lumens or by embolization of plaques. AID, when symptomatic, may present with a triad of claudication, impotence, and absence of femoral pulses, a triad also referred as Leriche Syndrome (LS).
View Article and Find Full Text PDFPort J Card Thorac Vasc Surg
November 2024
Angiologia e Cirurgia Vascular, ULS de São João, Porto, Portugal; UnIC@RISE, Department of Surgery and Physiology, Faculty of Medicine of the University of Porto, Porto, Portugal.
Port J Card Thorac Vasc Surg
January 2025
Cardiothoracic department, Santa Cruz Hospital, Portugal.
Ann Surg Oncol
January 2025
Department of Gynecologic Oncology, School of Medicine, Women's Hospital, Zhejiang University, Hangzhou, Zhejiang, China.
Background: This study aimed to explore the relationship of cervical tumor lesion location (CTLL) with bilateral parametrial involvement (PI) and pelvic lymph node metastasis (LNM).
Methods: The study retrospectively analyzed the clinicopathologic and imaging data of patients with cervical squamous cell carcinoma (SCC) retrieved from multiple centers. According to the CTLL, patients were allocated to three groups: a middle one third group, a unilaterally dominant group, and the entire-region group.
Sci Rep
January 2025
University of Ulsan, 93 Daehak-ro, Nam-gu, Ulsan, 680-749, Republic of Korea.
This study employed large eddy simulation (LES) with the wall-adapting local eddy-viscosity (WALE) model to investigate transitional flow characteristics in an idealized model of a healthy thoracic aorta. The OpenFOAM solver pimpleFoam was used to simulate blood flow as an incompressible Newtonian fluid, with the aortic walls treated as rigid boundaries. Simulations were conducted for 30 cardiac cycles and ensemble averaging was employed to ensure statistically reliable results.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!