The purpose of this study is to develop a method for use at extremely low-dose ranges and to decrease the uncertainty outside the recommended range of Gafchromic RTQA2 (RTQA2). By this method, the CT dose including the scattered radiation region can be grasped. The base density was increased by ultraviolet (UV)-ray preirradiation. RTQA2 was irradiated with UV-A rays for 26 and 40 h. Subsequently, RTQA2 was exposed to 2, 4, 6, 8, 10, 25, 50, 75, 100, 150, 200 and 250 mGy X-rays using a segmentation method. Calibration curves with and without UV-A irradiation were compared. The calibration curve with 40-h UV-A ray irradiation was the most linear, and a steeper slope area was not observed. The uncertainty in the calibration curve was reduced (p < 0.05). UV-A ray irradiation is an effective method for treating RTQA2; the accuracy in the extremely low-dose range of RTQA2 was improved.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1093/rpd/ncab115 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!