Purpose The purpose of this study is to compare optic disc dose (ODD) between I and Pd Collaborative Ocular Melanoma Study (COMS) plaques in ocular brachytherapy. Methods A previously validated in-house brachytherapy dose calculation program was used for ODD calculations. ODD was calculated as a function of tumor margin-to-optic disc distance (DT) up to 5 mm for various tumor basal dimensions (BDs), for a prescription depth of 5 mm, and for standard and notched COMS plaques loaded with I (model: IAI-125A) and Pd (model: IAPd-103A) seeds. ODD calculations were repeated for prescription depths from 2 mm to 10 mm in 1 mm intervals. A prescribed dose of 85 Gy (irradiation time: 120 hours) was normalized to each prescription depth. Dose conversion factors (DCFs) for each prescription depth were calculated by taking a ratio of [total reference air kerma (TRAK) per seed] to [TRAK per seed]. ODD reduction by notched COMS plaques was calculated for each prescription depth by subtracting ODD for notched COMS plaques from ODD for standard COMS plaques. Results Trends of ODD as a function of DT for various BDs are similar between the two seed types in both standard and notched COMS plaques. However, due to the energy difference, there exists a transition distance (d) for each BD in each plaque at which ODD for I COMS plaques equals that for Pd COMS plaques. For small BDs, at DT