AI Article Synopsis

  • The study aims to compare optic disc dose (ODD) between iodine (I) and palladium (Pd) seeds used in ocular brachytherapy for treating melanoma.
  • A validated dose calculation program was utilized to determine ODD based on various tumor sizes and optic disc distances, examining different plaque types and prescription depths.
  • Results show that ODD varies with tumor distance and size, revealing key differences between the two seed types, particularly in how notched plaques increase ODD reduction, especially for Pd seeds.

Article Abstract

Purpose The purpose of this study is to compare optic disc dose (ODD) between I and Pd Collaborative Ocular Melanoma Study (COMS) plaques in ocular brachytherapy. Methods A previously validated in-house brachytherapy dose calculation program was used for ODD calculations. ODD was calculated as a function of tumor margin-to-optic disc distance (DT) up to 5 mm for various tumor basal dimensions (BDs), for a prescription depth of 5 mm, and for standard and notched COMS plaques loaded with I (model: IAI-125A) and Pd (model: IAPd-103A) seeds. ODD calculations were repeated for prescription depths from 2 mm to 10 mm in 1 mm intervals. A prescribed dose of 85 Gy (irradiation time: 120 hours) was normalized to each prescription depth. Dose conversion factors (DCFs) for each prescription depth were calculated by taking a ratio of [total reference air kerma (TRAK) per seed]  to [TRAK per seed]. ODD reduction by notched COMS plaques was calculated for each prescription depth by subtracting ODD for notched COMS plaques from ODD for standard COMS plaques. Results Trends of ODD as a function of DT for various BDs are similar between the two seed types in both standard and notched COMS plaques. However, due to the energy difference, there exists a transition distance (d) for each BD in each plaque at which ODD for I COMS plaques equals that for Pd COMS plaques. For small BDs, at DT d, the opposite is observed. For the largest 1-3 BD(s), contrarily, d occurs within the tumor, and thus, ODD for I COMS plaquesis always higher. Trends of ODD reduction by notched COMS plaques as a function of DT for various BDs are the same for the two seed types except that maximum ODD reduction by Pd COMS notched plaques is larger. DCF increases with increasing prescription depth for both seed types. Conclusions There exist ODD differences between I and Pd COMS plaques and the differences depend on DT, BD, plaque size, and prescription depth.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8317005PMC
http://dx.doi.org/10.7759/cureus.15980DOI Listing

Publication Analysis

Top Keywords

coms plaques
44
prescription depth
24
notched coms
20
coms
13
plaques
12
odd reduction
12
seed types
12
odd
11
optic disc
8
disc dose
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!