Background And Objective: Estimates of parameters used to select patients for endovascular thrombectomy (EVT) for acute ischemic stroke differ among software packages for automated computed tomography (CT) perfusion analysis. To determine impact of these differences in decision making, we analyzed intra-observer and inter-observer agreement in recommendations about whether to perform EVT based on perfusion maps from 4 packages.

Methods: Perfusion CT datasets from 63 consecutive patients with suspected acute ischemic stroke were retrospectively postprocessed with 4 packages of Minerva, RAPID, Olea, and IntelliSpace Portal (ISP). We used Pearson correlation coefficients and Bland-Altman analysis to compare volumes of infarct core, penumbra, and mismatch calculated by Minerva and RAPID. We used kappa analysis to assess agreement among decisions of 3 radiologists about whether to recommend EVT based on maps generated by 4 packages.

Results: We found significant differences between using Minerva and RAPID to estimate penumbra (67.39±41.37mL vs. 78.35±45.38 mL, p < 0.001) and mismatch (48.41±32.03 vs. 61.27±32.73mL, p < 0.001), but not of infarct core (p = 0.230). Pearson correlation coefficients were 0.94 (95%CI:0.90-0.96) for infarct core, 0.87 (95%CI:0.79-0.91) for penumbra, and 0.72 (95%CI:0.57-0.83) for mismatch volumes (p < 0.001). Limits of agreements were (-21.22-25.02) for infarct core volumes, (-54.79-32.88) for penumbra volumes, and (-60.16-34.45) for mismatch volumes. Final agreement for EVT decision-making was substantial between Minerva vs. RAPID (k = 0.722), Minerva vs. Olea (k = 0.761), and RAPID vs. Olea (k = 0.782), but moderate for ISP vs. the other three.

Conclusions: Despite quantitative differences in estimates of infarct core, penumbra, and mismatch using 4 software packages, their impact on radiologists' decisions about EVT is relatively small.

Download full-text PDF

Source
http://dx.doi.org/10.3233/XST-210898DOI Listing

Publication Analysis

Top Keywords

acute ischemic
12
ischemic stroke
12
minerva rapid
12
software packages
8
evt based
8
higher agreement
4
agreement endovascular
4
endovascular treatment
4
treatment decision-making
4
decision-making parametric
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!