We report a unique case of delivery of inappropriate implantable cardioverter-defibrillator therapies related to a "perfect storm": presence of an integrated lead, insufficient lead slack related to right heart dilation resulting in shock coil misplacement, myocarditis with loss of R waves, and the concomitant occurrence of an incessant atrial tachycardia. ().
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8302787 | PMC |
http://dx.doi.org/10.1016/j.jaccas.2021.02.041 | DOI Listing |