Artifacts in contrast-enhanced mammography: are there differences between vendors?

Clin Imaging

University of Western Australia Medical School, Perth, Western Australia 6009, Australia; Department of Radiology, Royal Perth Hospital, Perth, Western Australia 6000, Australia. Electronic address:

Published: December 2021

Purpose: Contrast-Enhanced Mammography (CEM) produces a dual-energy subtracted (DES) image that demonstrates iodine uptake (neovascularity) in breast tissue. We aim to review a range of artifacts on DES images produced using equipment from two different vendors and compare their incidence and subjective severity.

Methods: We retrospectively reviewed CEM studies performed between September 2013 and March 2017 using GE Senographe Essential (n = 100) and Hologic Selenia Dimensions (n = 100) equipment. Artifacts were categorized and graded in severity by a subspecialist breast radiologist and one of two medical imaging technologists in consensus. The incidence of artifacts between vendors was compared by calculating the relative risk, and the severity gradings were compared using a Wilcoxon rank-sum test.

Results: Elephant rind, corrugations and the black line on chest wall artifact were seen exclusively in Hologic images. Artifacts such as cloudy fat, negative rim around lesion and white line on pectoral muscle were seen in significantly more Hologic images (p < 0.05) whilst halo, ripple, skin line enhancement, black line on pectoral muscle, bright pectorals, chest wall high-lighting and air gap were seen in significantly more GE images (p < 0.05). The severity gradings for cloudy fat had a significantly higher mean rank in Hologic images (p < 0.001) whilst halo and ripple artifacts had a significantly higher mean rank in GE images (p < 0.001 and p = 0.028 respectively).

Conclusion: The type, incidence and subjective severity of CEM-specific artifacts differ between vendors. Further research is needed, but differences in algorithms used to produce the DE image are postulated to be a significant contributor.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinimag.2021.06.031DOI Listing

Publication Analysis

Top Keywords

contrast-enhanced mammography
8
hologic images
8
artifacts
5
artifacts contrast-enhanced
4
mammography differences
4
differences vendors?
4
vendors? purpose
4
purpose contrast-enhanced
4
mammography cem
4
cem produces
4

Similar Publications

Neoadjuvant Therapy (NT) has become the gold standard for treating locally advanced Breast Cancer (BC). The assessment of pathological response (pR) post-NT plays a crucial role in predicting long-term survival, with Contrast-Enhanced Magnetic Resonance Imaging (MRI) currently recognised as the preferred imaging modality for its evaluation. Traditional imaging techniques, such as Digital Mammography (DM) and Ultrasonography (US), encounter difficulties in post-NT assessments due to breast density, lesion changes, fibrosis, and molecular patterns.

View Article and Find Full Text PDF

Contrast-enhanced mammography vs MRI: Which breast imaging modality leads in lesion detection?

Eur Radiol

December 2024

Department of Medical Imaging, Sackler School of Medicine, Sourasky Medical Center, Tel Aviv University, Tel Aviv, Israel.

View Article and Find Full Text PDF

Background: Under the Breast Imaging Reporting and Data System (BI-RADS), category 4 lesions have a high probability of malignancy. This study sought to investigate the efficacy of a model that combined the BI-RADS score with the enhancement score and clinical indicators in the diagnosis of BI-RADS 4 lesions based on contrast-enhanced spectral mammography (CESM) in breast cancer patients.

Methods: The data of female patients with BI-RADS scores of 4 who underwent CESM at the Department of Medical Imaging of the Third Affiliated Hospital of Soochow University from January 2018 to July 2023 were retrospectively collected.

View Article and Find Full Text PDF

In alloplastic breast reconstruction, the choice of implant positioning and the selection of periprosthetic devices is a critical and challenging decision. Surgeons must navigate between various biologic and synthetic meshes, including acellular dermal matrices (ADM). This study aimed to propose a simple selection tool for periprosthetic devices in prepectoral breast reconstruction.

View Article and Find Full Text PDF

Objectives: Asymmetries on screening contrast-enhanced mammography (CEM) often lead to patient recall. However, in diagnostic settings, negative CEM has effectively classified these as normal or benign, questioning the need for further workup of non-enhancing asymmetries (NEAs).

Material And Methods: A computational search of all screening CEM examinations performed between December-2012 and June-2021 was conducted to identify cases reporting NEAs.

View Article and Find Full Text PDF