Aim: To compare the differences and consistency of IOL-Master 700 biometers applying swept optical coherence tomography with the conventional IOL-Master 500 applying partial coherence interference in terms of the ocular biological parameters in adolescents with ametropia.
Methods: A total of 110 adolescents (110 eyes) with ametropia were collected, including 55 males and 55 females; age 10.69±2.81y. Ocular biological measurements were taken by IOL-Master 700 and IOL-Master 500 respectively to obtain biological parameters including axial length (AL), mean corneal anterior surface keratometry (K), anterior chamber depth (ACD), and horizontal corneal diameter (WTW). Paired -test was used to compare the differences between the two instruments. The intra-group correlation coefficient (ICC) and the Bland-Altman analysis were used to evaluate the consistency of parameter measurements between the two instruments for the four biological parameters.
Results: Statistical analysis showed that there was no significant difference in the K value measured by IOL-Master 700 and IOL-Master 500 (=-1.644, =0.116). The average differences of the AL, ACD, and WTW distances between the two instruments are 0.028, 0.101 and 0.064 mm respectively, and the differences are statistically significant (=2.644, 12.505, 3.911, <0.001). The consistency study results indicated high correlation in the measurement of AL, K, ACD and WTW between the two instruments (ICC=0.994, 0.873, 0.927, 0.912).
Conclusion: The novel biometric instrument IOL-Master 700 makes no difference with IOL-Master 500 in the measurement of K. There are some differences in the values of AL, ACD, and WTW. However, the two instruments show good consistency in these four biological measurements. The measured values of K are interchangeable between the instruments. These two types of biometrics can be used as mutual reference in consideration of that the differences in AL, ACD, and WTW measurements are not sufficient to produce clinically meaningful differences.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8243180 | PMC |
http://dx.doi.org/10.18240/ijo.2021.07.08 | DOI Listing |
J Cataract Refract Surg
January 2025
Yokohama Sky Building Eye Clinic, 2-19-12 Takashima, Yokohama, Nishi-ku, Kanagawa, 220-0011, Japan.
Purpose: To compare the refractive accuracy of the Barrett True axial length (BTAL) formula, newly integrated into ARGOS, with that of the Barrett Universal II (BUII) formula calculated using axial length (AL) from IOL Master 700.
Setting: Private clinics in Kanagawa, Japan.
Design: Retrospective observational study.
J Cataract Refract Surg
January 2025
Kepler University Hospital, University Clinic for Ophthalmology and Optometry.
Purpose: To comprehensively evaluate the tilt of the crystalline lens in phakic eyes and its relationship with optical biometry parameters.
Setting: The retrospective analysis was conducted at the Department for Ophthalmology and Optometry, Kepler University Clinic in Linz, Austria.
Design: Lens tilt assessment was performed through a combined analysis of optical biometry (IOL Master 700) and anterior segment optical coherence tomography (AS-OCT) data to understand its relationship.
Front Artif Intell
January 2025
School of Medicine, Stanford University, Palo Alto, CA, United States.
Given close relationships between ocular structure and ophthalmic disease, ocular biometry measurements (including axial length, lens thickness, anterior chamber depth, and keratometry values) may be leveraged as features in the prediction of eye diseases. However, ocular biometry measurements are often stored as PDFs rather than as structured data in electronic health records. Thus, time-consuming and laborious manual data entry is required for using biometry data as a disease predictor.
View Article and Find Full Text PDFSci Rep
December 2024
Department of Ophthalmology, Rajavithi Hospital and College of Medicine, Rangsit University, Bangkok, Thailand.
Relative anterior microphthalmos (RAM) is a rare ocular condition characterized by disproportionately small anterior segments but normal axial length (corneal diameter < 11 mm and axial length > 20 mm). This study aimed to determine the prevalence of RAM and its association with glaucoma utilizing IOL Master 700 data (Carl Zeiss Meditec, Jena, Germany). A retrospective analysis was conducted of the biometric parameters of 6,407 eyes, and 115 cases of RAM were identified.
View Article and Find Full Text PDFActa Ophthalmol
December 2024
Department of Ophthalmology and Vision Science, Eye and ENT Hospital, Fudan University, Shanghai, China.
Purpose: To evaluate the long-term effects of implantable collamer lens (ICL) V4c on the biomechanical and morphological properties of crystalline lenses in Chinese adults with high myopia using Brillouin microscopy (BM).
Methods: In this case-control study, 102 highly myopic eyes from 102 patients (18-48 years, preoperative spherical equivalent (SE): -6.00--21.