  • In UHPLC, high pressures lead to frictional heating that causes band broadening through radial diffusion and longitudinal dispersion, but this can be minimized with special techniques like vacuum jacketed tubing.
  • Utilizing vacuum jacketed column technology significantly improved metabolite profiling in urine, resulting in narrower peak widths and lower peak tailing compared to conventional LC/MS.
  • The enhanced performance of the VJC system not only increased the peak capacity and the number of detected features but also improved the intensity of mass spectrometry (MS) peaks, leading to better data quality and detection limits.

Article Abstract

In UHPLC, frictional heating from the eluent flowing through the column at pressures of ca. 10-15 Kpsi causes radial diffusion via temperature differences between the center of the column and its walls. Longitudinal dispersion also occurs due to temperature gradients between the inlet and outlet. These effects cause band broadening but can be mitigated via a combination of vacuum jacketed stainless steel tubing, reduced column end nut mass, and a constant temperature in the column from heating the inlet fitting. Here, vacuum jacketed column (VJC) technology, employing a novel column housing located on the source of the mass spectrometer and minimized tubing from the column outlet to the electrospray probe, was applied to profiling metabolites in urine. For a 75 s reversed-phase gradient separation, the average peak widths for endogenous compounds in urine were 1.2 and 0.6 s for conventional LC/MS and VJC systems, respectively. The peak tailing factor was reduced from 1.25 to 1.13 when using the VJC system compared to conventional UHPLC, and the peak capacity increased from 65 to 120, with a 25% increase in features detected in urine. The increased resolving power of the VJC system reduced co-elution, simplifying MS and MS/MS spectra, providing a more confident metabolite identification. The increased LC performance also gave more intense MS peaks, with a 10-120% increase in response, improving the quality of the MS data and detection limits. Reducing the LC gradient duration to 37 s gave peak widths of ca. 0.4 s and a peak capacity of 84.

Download full-text PDF

Source
http://dx.doi.org/10.1021/acs.analchem.1c01982DOI Listing

Publication Analysis

Top Keywords

vacuum jacketed
12
column
8
jacketed column
8
peak widths
8
vjc system
8
peak capacity
8
peak
5
high-throughput uhplc/ms/ms-based
4
uhplc/ms/ms-based metabolic
4
metabolic profiling
4