Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objective: To evaluate the impact of an electronic witnessing system (EWS) on witnessing standard operating procedures and to assess embryologist perceptions of the EWS.
Design: Prospective cohort study.
Setting: Private in vitro fertilization laboratory network.
Patients: None.
Interventions: None.
Main Outcome Measures: The time difference between manual and electronic double-witnessing procedures, and embryologist perceptions of the EWS.
Results: From 342 witnessing times analyzed (114 EWS, 114 manual, and 114 interruptions to witnesses), the EWS reduced mean (SD) total witnessing time (in seconds) by 91.5 (23.6) for intracytoplasmic sperm injection, 62.0 (17.9) for Day 3 embryo assessment, 58.3 (18.9) for fresh embryo transfer, and 59.4 (13.3) for frozen embryo transfer. This time reduction significantly decreased the overall time required for double-witnessing by 3.1- to 5.2-fold. A survey with 50 embryologists within the laboratory network indicated that most embryologists considered the EWS to improve sample traceability (78.3%), reduce errors in labeling issues (80.4%), and reduce the risk of sample mismatch errors by minimizing disruptions (60.9%). Furthermore, 82.6% thought that visual completion of the EWS dashboard provided peace of mind when leaving work and 84.8% were more confident knowing that all procedures were completed according to the EWS.
Conclusions: An EWS can improve laboratory efficiency by significantly decreasing the time required for witnessing procedures and by minimizing interruptions. The EWS was well perceived by embryologists and laboratory managers and enhanced their confidence and peace of mind with regard to witnessing compliance and safety/accuracy.