Background: Although electrical impedance tomography (EIT) is widely used for monitoring regional ventilation distribution, reference values have yet to be established for clinical use. The present study aimed to evaluate the feasibility of creating reference values for standard EIT parameters for potential clinical application.

Methods: A total of 75 participants with healthy lungs were included in this prospective study (male:female, 48:27; age, 34±14 years; height, 172±7 cm; weight, 73±12 kg). The subjects were examined during spontaneous breathing in the supine position. EIT measurements were performed at the level of the 4 intercostal space. Commonly used EIT-based parameters, including the center of ventilation (CoV), dorsal and most dorsal fractions of ventilation distribution (TV and TV respectively), global inhomogeneity (GI) index, and standard deviation of regional ventilation delay index (RVD) were calculated.

Results: Following outlier detection, EIT data from 71 subjects were finally evaluated. The values of the evaluated parameters were: CoV, 48.7%±1.7%; TVD, 48.1%±5.4%; TV, 7.1%±1.8%; GI, 0.49±0.04; and RVD, 7.0±2.0. The coefficients of variation for CoV and GI were low (0.03 and 0.07, respectively), but those for TV and RVD were comparatively high (0.26 and 0.28, respectively). None of the evaluated parameters showed a significant correlation with age. The GI index showed a weak but significant correlation with body mass index (R=0.29, P=0.01). The RVD was slightly higher in males than in females.

Conclusions: Our study indicated that CoV and GI were stable parameters with small coefficients of variation in participants with healthy lungs. The creation of EIT parameter reference values for setting treatment targets may be feasible.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8246208PMC
http://dx.doi.org/10.21037/atm-20-7442DOI Listing

Publication Analysis

Top Keywords

reference values
16
regional ventilation
12
ventilation distribution
12
healthy lungs
12
values established
8
electrical impedance
8
impedance tomography
8
participants healthy
8
evaluated parameters
8
coefficients variation
8

Similar Publications

Purpose: The present study examined masseteric vestibular evoked myogenic potential (mVEMP) responses in children and compared these results with those of adults, using both air- and bone-conduction modes of stimulation.

Method: Fifteen children and fifteen adults with hearing thresholds below 15 dB HL were considered for the study. Ipsilateral 500-Hz narrow-band chirp (NB CE-chirp) evoked mVEMP responses were elicited via a zygomatic montage through both air- and bone-conduction modes of stimulation.

View Article and Find Full Text PDF

The emergence of toothpastes containing different abrasive and whitening substances has been a constant concern among dental professionals. The aim of the present study was to perform an in vitro assessment of the surface topography of nanoparticle composite resins subjected to simulated brushing with dentifrices. Test samples were prepared with Filtek Universal (3M ESPE), Filtek Bulkfill (3M ESPE) and Z350 (3M ESPE), with 24 samples per resin.

View Article and Find Full Text PDF

Although it is recognized that periodontal disease negatively impacts quality of life, there is no validated instrument to assess this impact in Brazil. This study aimed to translate, cross-culturally adapt, and validate the OHIP 14 PD (Oral Health Impact Profile Applied to Periodontal Diseases) for application among Brazilian patients. The original instrument was translated and validated into Brazilian Portuguese in a cross-sectional study with 110 participants recruited from a Dental School clinic.

View Article and Find Full Text PDF

New spirometry recommendations from the Brazilian Thoracic Association - 2024 update.

J Bras Pneumol

January 2025

. Laboratório de Função Pulmonar, Instituto de Assistência ao Servidor Público Estadual de São Paulo Francisco Morato Pereira - IAMSPE-FMO - São Paulo (SP), Brasil.

The latest pulmonary function guideline from the Brazilian Thoracic Association was published in 2002, since which there have been updates to international guidelines (mainly those from the European Respiratory Society and the American Thoracic Society), as well as new national and international publications on various aspects of the performance, interpretation, and clinical implications of spirometry. Despite those updates, a careful analysis of what applies to the reality in Brazil is essential, because there have been studies that evaluated individuals who are representative of our population and who could show responses different from those of individuals in other regions of the world. This document is the result of the work of a group of specialists in pulmonary function who evaluated relevant scientific articles that could be applicable to the population of Brazil.

View Article and Find Full Text PDF

Serum Nitric Oxide, Endothelin-1 Correlates Post-Procedural Major Adverse Cardiovascular Events among Patients with Acute STEMI.

Arq Bras Cardiol

January 2025

Department of Cardiovascular Medicine - Shengzhou People's Hospital (Shengzhou Branch of the First Affiliated Hospital of Zhejiang University School of Medicine, the Shengzhou Hospital of Shaoxing University), Zhejiang - China.

Background: ST-segment elevation myocardial infarction (STEMI) is a common and severe form of acute myocardial infarction (AMI).

Objectives: The study aimed to investigate the relationship between serum nitric oxide (NO) and endothelin-1 (ET-1) levels with the severity of STEMI and their predictive value for major adverse cardiovascular events (MACE) within one year after percutaneous coronary intervention (PCI) in STEMI patients.

Methods: The retrospective study was conducted on 269 STEMI patients who underwent PCI.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!