Validity of the Handheld Expiratory Flowmeter for COPD Screening in the Primary Care Setting of China.

Int J Chron Obstruct Pulmon Dis

State Key Laboratory of Respiratory Disease, Guangzhou Institute of Respiratory Diseases, National Clinical Research Center for Respiratory Disease Guangzhou, First Affiliated Hospital of Guangzhou Medical University, Guangzhou Medical University, Guangzhou, Guangdong Province, People's Republic of China.

Published: August 2021

Purpose: The use of simple and affordable screening tools for chronic obstructive pulmonary disease (COPD) is limited. We aimed to assess the validity of a handheld expiratory flowmeter (Vitalograph Ltd., COPD-6, Ireland) for COPD screening in Chinese primary care settings.

Methods: In our cross-sectional study, subjects were randomly selected in eight primary care settings. Tests with the handheld expiratory flowmeter and the conventional spirometry were sequentially performed on all participants. The correlation between the handheld expiratory flowmeter and the conventional spirometry was determined. Validity was determined by the area under the receiver operator characteristic curve (AUC) of the forced expiratory volume in one second (FEV)forced expiratory volume in six seconds (FEV) that used to detect airway obstruction. The sensitivity, specificity, predictive values, and likelihood ratio were calculated according to different FEV/FEV cut-off points.

Results: A total of 229 subjects (15.4%) were diagnosed with airflow limitation by conventional spirometry. FEV, FEV, and FEV/FEV measured by the handheld expiratory flowmeter were correlated with FEV, FVC, and FEV/FVC measured by the conventional spirometry (=0.889, 0.835 and 0.647, p<0.001), respectively. AUC of the FEV/FEV to determine airflow obstruction was 0.857 (95% CI: 0.826 to 0.888). No significant difference of AUC was observed between the symptomatic group and the asymptomatic group (AUC=0.869 vs 0.843, P=0.425). A similar phenomenon was found in the AUC of smokers and never-smokers (AUC=0.862 vs 0.840; P=0.515). The cut-off point for FEV/FEV was 0.77 and the corresponding sensitivity and specificity were 71.2% and 89.8%, respectively.

Conclusion: The handheld expiratory flowmeter might be used as a screening device for COPD in Chinese primary care settings.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8275149PMC
http://dx.doi.org/10.2147/COPD.S312190DOI Listing

Publication Analysis

Top Keywords

handheld expiratory
20
expiratory flowmeter
20
conventional spirometry
16
primary care
12
validity handheld
8
copd screening
8
flowmeter conventional
8
expiratory volume
8
expiratory
7
flowmeter
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!