Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: The increased use of ionizing radiation for diagnostic purpose has resulted in an increase in the world population dose. Patient dosimetry in X-ray diagnostic radiology is required to establish diagnostic reference levels (DRLs) and to assess the average dose received by organs and tissues. International bodies have recommended DRLs to be based on dosimetric quantities.
Aim: To cater to the increased requirement for dosimetry in diagnostic radiology, international guidelines are provided to establish and disseminate traceable calibration for dosimeters used in X-ray diagnostic radiology. X-ray diagnostic beams established are standardized using a diagnostic range free-air ionization chamber (DFAIC) (20-150 kV).
Materials And Methods: Characterization of the DFAIC and determination of the correction factors for the air kerma measurements were evaluated experimentally and by theoretical calculations.
Results: The paper details the establishment of 18 diagnostic beam qualities using DFAIC along with the associated uncertainties. The overall uncertainty for the air kerma measurements was within ±0.5% at 1 sigma level. Eight diagnostic range air kerma measurements using DFAIC were compared with the medium energy primary standard FAIC (50-300 kV) maintained in the laboratory.
Conclusion: The air kerma rates agreed within ±1% and are within the overall standard uncertainty of both the chambers at the time of the comparison. Dissemination to the users in the field of diagnostic radiology in the country has been carried out by calibrating their ionization chambers and solid-state detector-based instruments against the DFAIC. The methodology followed to standardize the beams using DFAIC and calibration of dosimeters is presented in this work.