Objectives: In some patients treated for urinary or fecal incontinence with sacral neuromodulation (SNM) persistence of symptoms, a reduction in efficacy or adverse effects of stimulation can occur. In such situations, further programming of the SNM device can help resolve problems. Infrequently hardware failure is detected. This article aims to provide practical guidance to solve sub-optimal outcomes (troubleshooting) occurring in the course of SNM therapy.

Materials And Methods: A systematic literature review was performed. Collective clinical experience from an expert multidisciplinary group was used to form opinion where evidence was lacking.

Results: Circumstances in which reprogramming is required are described. Actions to undertake include changes of electrode configuration, stimulation amplitude, pulse frequency, and pulse width. Guidance in case of loss of efficacy and adverse effects of stimulation, developed by a group of European experts, is presented. In addition, various hardware failure scenarios and their management are described.

Conclusions: Reprogramming aims to further improve patient symptoms or ensure a comfortable delivery of the therapy. Initial changes of electrode configuration and adjustment of stimulation parameters can be performed at home to avoid unnecessary hospital visits. A logical and stepwise approach to reprogramming can improve the outcome of therapy and restore patient satisfaction.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9291141PMC
http://dx.doi.org/10.1111/ner.13494DOI Listing

Publication Analysis

Top Keywords

sacral neuromodulation
8
sub-optimal outcomes
8
efficacy adverse
8
adverse effects
8
effects stimulation
8
hardware failure
8
changes electrode
8
electrode configuration
8
reprogramming
4
reprogramming sacral
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!