Purpose: To verify the practicality and safety of a treatment chair with six degrees of freedom (6DTC) through demonstrating the efficacy of the workflow in clinical settings and analyzing the obtained technical data, including intra-fraction patient movement during the use of the 6DTC.

Materials And Methods: A clinical study was designed and conducted to test the clinical treatment workflow and the safety of the 6DTC. Based on the demonstrated dosimetric advantages, fifteen patients with head and neck tumors were selected and treated with the 6DTC. The positional error at the first beam position (PE-B1) and the second beam position (PE-B2) were analyzed and compared with the results from daily quality assurance (QA) procedures of the 6DTC and imaging system performed each day before clinical treatment. The intra-fraction patient movement was derived from the total patient alignment positional error and the QA data based on a Gaussian distribution formulism.

Results: The QA results showed sub-millimeter mechanical accuracy of the 6DTC over the course of the clinical study. For 150 patient treatment fractions, the mean deviations between PE-B1 and PE-B2 were 0.13mm (SD 0.88mm), 0.25mm (SD 1.17mm), -0.57mm (SD 0.85mm), 0.02° (SD 0.35°), 0.00° (SD 0.37°), and -0.02° (SD 0.37°) in the x, y, z (translational), and u, v, w (rotational) directions, respectively. The calculated intra-fraction patient movement was -0.08mm (SD 0.56mm), 0.71mm (SD 1.12mm), -0.52mm (SD 0.84mm), 0.10° (SD 0.32°), 0.09° (SD 0.36°), and -0.04° (SD 0.36°) in the x, y, z, u, v, w directions, respectively.

Conclusions: The performance stability of the 6DTC was satisfactory. The position accuracy and intra-fraction patient movement in an upright posture with the 6DTC were verified and found adequate for clinical implementation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8260974PMC
http://dx.doi.org/10.3389/fonc.2021.694749DOI Listing

Publication Analysis

Top Keywords

intra-fraction patient
16
patient movement
16
clinical implementation
8
treatment chair
8
clinical study
8
clinical treatment
8
positional error
8
beam position
8
clinical
7
6dtc
7

Similar Publications

Background/objectives: An on-board imager on a linear accelerator allows the acquisition of kV-2D images during irradiation. Overlaying specific structures on these images enables the visual verification of movement at regular frequencies. Our aim was to validate this tracking method for the stereotactic treatment of bone metastases.

View Article and Find Full Text PDF

Background: Immobilization devices are essential for maintaining accurate and repeatable patient positioning in radiotherapy. This study aimed to evaluate the setup errors and dosimetric deviation induced by the deformation of immobilization devices in thoracic cancer radiotherapy using CT-linac.

Materials And Methods: A retrospective analysis was conducted on 40 thoracic cancer patients who underwent radiotherapy, using vacuum cushion (VC) and thermoplastic mask (TM) for immobilization.

View Article and Find Full Text PDF

Objective: To evaluate the intra-fraction and inter-fraction positional deviations in head and neck tumor patients undergoing intensity-modulated radiation therapy (IMRT) guided by cone-beam CT (CBCT), as well as the correction capability and stability of the HexaPOD evo RT 6D couch in addressing these deviations.

Methods: From May 2019 to April 2022, 59 consecutive patients with head and neck tumors were enrolled at the Department of Radiation Oncology, Peking University Third Hospital.Using the Elekta AXESSE image-guided stereotactic treatment system, a pre-treatment CBCT scan was performed, followed by bone window mode registration with the planning reference images.

View Article and Find Full Text PDF

Linac-based stereotactic radiosurgery (SRS) with planning target volume (PTV) margins <1 mm has become increasingly common in recent years. Optical surface imaging for surface-guided radiation therapy (SGRT) is often used for intra-fraction motion monitoring during these treatments to facilitate the use of a smaller PTV margin by providing real-time quantitative patient positioning information. However, rotating the couch introduces errors to SGRT-reported translations and rotations that can be problematic for SRS treatments with non-coplanar arcs and very small PTV margins.

View Article and Find Full Text PDF

A systematic review of prostate bed motion and anisotropic margins in post-prostatectomy external beam radiotherapy.

Tech Innov Patient Support Radiat Oncol

December 2024

Applied Radiation Therapy Trinity (ARTT), Discipline of Radiation Therapy, School of Medicine, Trinity St. James's Cancer Institute, Trinity College Dublin, University of Dublin, Dublin, Ireland.

Background: Prostate bed (PB) motion may lead to geographical miss of the target volume in post-prostatectomy radiotherapy (RT). Optimal clinical target volume (CTV) to planning target volume (PTV) margins prevent geographical miss and unnecessary irradiation of normal tissue. There is little data available informing appropriate CTV to PTV margins in the post-prostatectomy setting.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!