Upgrade and absolute calibration of the JET scintillator-based fast-ion loss detector.

Rev Sci Instrum

Departamento de Ingenieria Mecanica y Fabricacion, Universidad de Sevilla, Camino de los Descubrimientos s/n, 41092 Sevilla, SpainCentro Nacional de Aceleradores (CNA) (Universidad de Sevilla, CSIC, Junta de Andalucia), Seville, SpainInstitute of Plasma Physics and Laser Microfusion, Hery 23, 01-497 Warsaw, PolandDepartamento de Fisica Atomica, Molecular y Nuclear, Universidad de Sevilla, 41012 Sevilla, SpainCCFE, Culham Science Centre, Abingdon, Oxon OX14 3DB, United KingdomOEAW, Institute for Theoretical Physics, University of Innsbruck, Innsbruck, AustriaLaboratory for Ion beam Interactions, Ruder Boskovic Institute, Zagreb, Croatia.

Published: April 2021

The JET FILD is a scintillator-based Fast-ion Loss Detector optimized to measure fusion-born alpha-particle losses. This work covers its upgrade and absolute calibration in preparation for the following JET DT experiments. A fast scintillator material (TG-Green) has been installed in the JET FILD. A heater jacket is installed around the fiber bundle, responsible for transmitting the light from the scintillator plate, to anneal the fiber obscuring due to neutron damage. The JET FILD has been upgraded with a 1 Mpx camera and 2 MHz photomultiplier data acquisition hardware. Full-orbit simulations give an estimate of the shading effects on the scintillator plate of the first wall structures and provide a synthetic signal of the JET FILD. A detector instrument function enables absolute values of fast-ion losses using calibration factors. The calibration factors are made available in a shot-to-shot basis for the characterized species and energies and with corrections for the diagnostic conditions. The fast acquisition system sets the Nyquist frequency (1 MHz) above the typical mode frequencies (≈10 kHz), thus making it possible to identify MHD-induced fast-ion losses.

Download full-text PDF

Source
http://dx.doi.org/10.1063/5.0043760DOI Listing

Publication Analysis

Top Keywords

jet fild
16
upgrade absolute
8
absolute calibration
8
scintillator-based fast-ion
8
fast-ion loss
8
loss detector
8
scintillator plate
8
fast-ion losses
8
calibration factors
8
jet
6

Similar Publications