A Low-Cost, Point-of-Care Test for Confirmation of Nasogastric Tube Placement via Magnetic Field Tracking.

Sensors (Basel)

School of Mechanical and Aerospace Engineering, Nanyang Technological University, Singapore 639798, Singapore.

Published: June 2021

In this work, we aim to achieve low-cost real-time tracking for nasogastric tube (NGT) insertion by using a tracking method based on two magnetic sensors. Currently, some electromagnetic (EM) tracking systems used to detect the misinsertion of the NGT are commercially available. While the EM tracking systems can be advantageous over the other conventional methods to confirm the NGT position, their high costs are a factor hindering such systems from wider acceptance in the clinical community. In our approach, a pair of magnetic sensors are used to estimate the location of a permanent magnet embedded at the tip of the NGT. As the cost of the magnet and magnetic sensors is low, the total cost of the system can be less than one-tenth of that of the EM tracking systems. The experimental results exhibited that tracking can be achieved with a root mean square error (RMSE) of 2-5 mm and indicated a great potential for use as a point-of-care test for NGT insertion, to avoid misplacement into the lung and ensure correct placement in the stomach.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8271631PMC
http://dx.doi.org/10.3390/s21134491DOI Listing

Publication Analysis

Top Keywords

magnetic sensors
12
tracking systems
12
point-of-care test
8
nasogastric tube
8
ngt insertion
8
tracking
7
ngt
5
low-cost point-of-care
4
test confirmation
4
confirmation nasogastric
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!