Intra-saccular device modeling for treatment planning of intracranial aneurysms: from morphology to hemodynamics.

Int J Comput Assist Radiol Surg

Instituto Pladema - CONICET, Universidad Nacional del Centro de la Provincia de Buenos Aires, Tandil, Argentina.

Published: October 2021

Motivation: Intra-saccular devices (ID), developed for the treatment of bifurcation aneurysms, offer new alternatives for treating complex terminal and bifurcation aneurysms. In this work, a complete workflow going from medical images to post-treatment CFD analysis is described and used in the assessment of a concrete clinical problem.

Materials And Methods: Two different intra-saccular device sizes were virtually implanted in 3D models of the patient vasculature using the ID-Fit method. After deployment, the local porosity at the closed end of the device in contact with the blood flow was computed. This porosity was then used to produce a CFD porous medium model of the device. Velocities and wall shear stress were assessed for each model.

Results: Six patients treated with intra-saccular devices were included in this work. For each case, 2 different device sizes were virtually implanted and 3 CFD simulations were performed: after deployment simulation with each size and before deployment simulation (untreated). A visible reduction in velocities was observed after device implantation. Velocity and WSS reduction was statistically significant (K-S statistics, [Formula: see text]).

Conclusions: Placement of different device size can lead to a partial filling of the aneurysm, either at the dome or at the neck, depending on the particular positioning by the interventionist. The methodology used in this work can have a strong clinical impact, since it provides additional information in the process of device selection using preoperative data.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s11548-021-02427-9DOI Listing

Publication Analysis

Top Keywords

intra-saccular device
8
intra-saccular devices
8
bifurcation aneurysms
8
device sizes
8
sizes virtually
8
virtually implanted
8
deployment simulation
8
device
7
intra-saccular
4
device modeling
4

Similar Publications

Quantitative radial force measurements of Woven EndoBridge devices.

Interv Neuroradiol

October 2023

Interventional MR Clinical R&D Institute, Ankara University, Ankara, Turkey.

Background: Lateral/radial forces and the mechanical properties of Woven EndoBridge (WEB) devices have significant importance for therapeutic success. In other words, adequate apposition of the lateral wall of a cerebral aneurysm is critical for preventing recurrence or re-rupture risk.

Objective: This study aimed to investigate the pressure values applied by different WEB devices to the lateral walls of aneurysms and the relationships between these pressure measurements and the diameters of WEB devices.

View Article and Find Full Text PDF
Article Synopsis
  • The WEB-IT study assessed the safety and effectiveness of the WEB device in treating wide neck bifurcation aneurysms (WNBAs) over a 5-year period with data from 150 patients.
  • Out of the patients, 83 had 5-year imaging follow-up, showing zero rebleeding events and a significant proportion achieving complete (58.1%) or adequate (87.2%) occlusion of their aneurysms.
  • The study concluded that the WEB device demonstrated long-term safety and efficacy, with durable occlusion rates and minimal need for retreatment, indicating it is a viable option for managing WNBAs.
View Article and Find Full Text PDF

Background And Purpose: Treating aneurysms with intra-saccular flow disruption is a feasible alternative to coil-embolization. Besides the established WEB device, the novel Contour Neurovascular System has emerged as a potentially easier alternative regarding sizing and deployment. We report the learning curve experienced at our center from the first 48 patients treated with Contour and compared it with 48 consecutive WEB cases.

View Article and Find Full Text PDF

Introduction: The Woven EndoBridge device (WEB) was developed as an alternative to treat Wide-Necked bifurcation aneurysms. It has proven to be effective and safe, however, cases of recanalization have been reported. The purpose of this study was to evaluate and quantify hemodynamic parameters and indexes with CFD of the intracranial aneurysms before and after WEB simulation and to establish their relationship to complete occlusion.

View Article and Find Full Text PDF

Simulation of intra-saccular devices for pre-operative device size selection: Method and validation for sizing and porosity simulation.

Comput Biol Med

August 2022

Instituto PLADEMA - CONICET, Universidad Nacional del Centro de la Provincia de Buenos Aires, Tandil, Argentina; Galgo Medical S.L., Barcelona, Spain.

Intra-saccular devices (ID) are novel braided devices used for complex intracranial aneurysms treatment. Treatment success is associated with correct device size selection. A technique that predicts the ID size within the aneurysm before intervention will provide a powerful computational tool to aid the interventionist during device selection.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!