Background: Cheese-wiring, the suture that cuts through the meniscus, is a well-known issue in meniscal repair. So far, contributing factors are neither fully understood nor sufficiently studied.
Hypothesis/purpose: To investigate whether the construct stiffness of repair sutures and devices correlates with suture cut-through (cheese-wiring) during load-to-failure testing.
Study Design: Controlled laboratory study.
Methods: In 131 porcine menisci, longitudinal bucket-handle tears were repaired using either inside-out sutures (n = 66; No. 0 Ultrabraid, 2-0 Orthocord, 2-0 FiberWire, and 2-0 Ethibond) or all-inside devices (n = 65; FastFix360, Omnispan, and Meniscal Cinch). After cyclic loading, load-to-failure testing was performed. The mode of failure and construct stiffness were recorded. A receiver operating characteristic curve analysis was performed to define the optimal stiffness threshold for predicting meniscal repair failure by cheese-wiring. The 2-tailed test and analysis of variance were used to test significance.
Results: Loss of suture fixation was the most common mode of failure in all specimens (58%), except for the Omnispan, which failed most commonly because of anchor pull-through. The Omnispan demonstrated the highest construct stiffness (30.8 ± 3.5 N/mm), whereas the Meniscal Cinch (18.0 ± 8.8 N/mm) and Ethibond (19.4 ± 7.8 N/mm) demonstrated the lowest construct stiffness. The Omnispan showed significantly higher stiffness compared with the Meniscal Cinch ( < .001) and Ethibond ( = .02), whereas the stiffness of the Meniscal Cinch was significantly lower compared with that of the FiberWire ( = .01), Ultrabraid ( = .04), and FastFix360 ( = .03). While meniscal repair with a high construct stiffness more often failed by cheese-wiring, meniscal repair with a lower stiffness failed by loss of suture fixation, knot slippage, or anchor pull-through. Meniscal repair with a stiffness >26.5 N/mm had a 3.6 times higher risk of failure due to cheese-wiring during load-to-failure testing (95% CI, 1.4-8.2; < .0001).
Conclusion: Meniscal repair using inside-out sutures and all-inside devices with a higher construct stiffness (>26.5 N/mm) was more likely to fail through suture cut-through (cheese-wiring) than that with a lower stiffness (≤26.5 N/mm).
Clinical Relevance: This is the first study investigating the impact of construct stiffness on meniscal repair failure by suture cut-through (cheese-wiring).
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8209834 | PMC |
http://dx.doi.org/10.1177/23259671211015674 | DOI Listing |
J Anat
January 2025
Institute of Sports Medicine Copenhagen, Department of Orthopedic Surgery, Copenhagen University Hospital - Bispebjerg-Frederiksberg, Copenhagen, Denmark.
Tendon injuries and disorders associated with mechanical tendon overuse are common musculoskeletal problems. Even though tendons play a central role in human movement, the intrinsic healing process of tendon is very slow. So far, it is known that tendon cell activity is supported by several interstitial cells within the tendon.
View Article and Find Full Text PDFArch Orthop Trauma Surg
January 2025
Department of Orthopaedics, Wright State University, 30 E Apple St., Suite 2200, Dayton, OH, 45409, USA.
Introduction: We propose and assess the biomechanical stability of medial column screw supplementation in a synthetic distal femur fracture model.
Materials And Methods: Twenty-four low density synthetic femora modeling osteoporotic, intraarticular distal femur fractures with medial metaphyseal comminution were split into two fixation groups: (1) lateral locking distal femur plate (PA- plate alone) and (2) lateral locking distal femur plate with a 6.5 mm fully threaded medial cannulated screw (PWS- plate with screw).
Front Robot AI
January 2025
School of Engineering, Institute of Science Tokyo, Tokyo, Japan.
Animal muscles have complex, three-dimensional structures with fibers oriented in various directions. The tongue, in particular, features a highly intricate muscular system composed of four intrinsic muscles and several types of extrinsic muscles, enabling flexible and diverse movements essential for feeding, swallowing, and speech production. Replicating these structures could lead to the development of multifunctional manipulators and advanced platforms for studying muscle-motion relationships.
View Article and Find Full Text PDFAngiology
January 2025
Family Center for Research with Ubuntu, Department of Paediatrics and Child Health, Stellenbosch University, Stellenbosch, South Africa.
Atherosclerosis often starts in childhood, tracking to adulthood. In children, early vascular disease can be detected as arterial stiffness. Carotid-femoral pulse wave velocity is considered the non-invasive gold standard method for measuring arterial stiffness and widely accepted for use in children.
View Article and Find Full Text PDFSci Rep
January 2025
Department of Mechanical Engineering, College of Engineering, King Khalid University, Abha, 61421, Saudi Arabia.
Fault ruptures induced by earthquakes pose a significant threat to constructions, particularly underground structures such as pile foundations. Among various foundation types, batter pile foundations are widely used due to their ability to resist inclined forces. To gain new insights into the response of batter pile groups to fault ruptures caused by earthquakes, this study investigates the deformation and failure mechanisms of batter pile groups due to the propagation of normal and reverse fault ruptures using 3D numerical modeling.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!