Background: Active detection of SARS-CoV-2 infection through testing is elementary for the control of COVID-19 pandemic. The implementation of large-scale RT-PCR testing has led to a rise in the demand for testing kits whose availability is always a concern.

Objective: To find out the feasibility of pooled testing in a high-throughput platform.

Methodology: Pooled testing was conducted in Roche cobas 6800 in 2 methods. Firstly, the simple two-stage testing algorithm was conducted for 1410 samples individually and then as pooled samples. Secondly, we evaluated the sensitivity of cobas 6800 for the detection of a single positive sample within a pool of negative samples.

Results: Implementing the five-sample Dorfman pooling to test 1410 samples, we identified 42 (2.9%) individual SARS-CoV-2-positive samples and 27 (9.5%) positive pool samples. The pooling strategy precisely identified all the positive samples. All individually negative samples were also accurately determined by pooling. There was 100% sensitivity of detecting positive samples in a pool of negative samples even up to 1:64 dilution. There was a threefold increase in total throughput in one-third of the cost per day.

Conclusion: A high-throughput platform such as Cobas 6800 can effectively increase the testing capacity by twofold to threefold by adopting the pooled testing strategy for successful management of SARS-CoV-2 and helping in the containment of community transmission.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8274989PMC
http://dx.doi.org/10.1002/jcla.23835DOI Listing

Publication Analysis

Top Keywords

pooled testing
16
cobas 6800
12
samples
9
sars-cov-2 infection
8
high-throughput platform
8
testing
8
1410 samples
8
samples individually
8
pool negative
8
positive samples
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!