Purifying air over the operating field with a new mobile laminar airflow device to reduce the possibility of airborne contamination.

J Cataract Refract Surg

From the Cincinnati Eye Institute , Cincinnati, Ohio (Osher, Figueiredo); University of Cincinnati, Cincinnati, Ohio (Osher); D'Olhos Hospital Dia (Figueiredo), S. J. Rio Preto, Brazil; Columbia University Harkness Eye Institute, New York, New York (Schneider); Hudson Eye Physicians and Surgeons, Jersey City, New Jersey (Schneider); Narvik Eye Institue, Narvik, Norway (Kratholm).

Published: October 2021

Purpose: To investigate the effectiveness of a mobile laminar airflow (LAF) device designed to reduce both airborne particles and lint fibers within the sterile field during cataract surgery.

Setting: Cincinnati Eye Institute, Cincinnati, Ohio.

Design: Prospective case series.

Methods: The number of airborne particles sized 0.5 μm, 1.0 μm, and 1.5 μm were measured at different locations and times during 116 routine phacoemulsification procedures with and without the use of a mobile LAF device. In a companion study, the presence of lint fibers in 99 eyes undergoing cataract surgery with a mobile LAF device was compared with 50 eyes in the control group. Lint fibers were differentiated as either falling onto the ocular surface or being carried into the sterile field by an instrument.

Results: A statistically significant reduction (P < .0005) occurred in all 3 particle sizes measurements when the LAF device was used: 79.0% (93.6 ± 16.1 vs 445.3 ± 30.9), 81.7% (11.30 ± 1.98 vs 61.85 ± 5.34), and 90.8% (0.241 ± 0.056 vs 2.624 ± 0.362) in the 0.5 μm, 1.0 μm and 5.0 μm particles, respectively. Lint fibers were identified in 18% (9/50) of eyes in the control group and 16.16% (16/99) of eyes in the LAF group. Although the number of lint fibers carried into the sterile field was similar in each group, the incidence of lint fibers falling onto the sterile field was reduced from 6% (3/50) to 0% (0/99) when the LAF was used (P = .014).

Conclusions: The mobile LAF device was highly effective in reducing the number of particulate matter and lint fibers within the sterile surgical field when used during cataract surgery.

Download full-text PDF

Source
http://dx.doi.org/10.1097/j.jcrs.0000000000000613DOI Listing

Publication Analysis

Top Keywords

lint fibers
28
laf device
20
sterile field
16
μm μm
16
mobile laf
12
mobile laminar
8
laminar airflow
8
airborne particles
8
particles lint
8
fibers sterile
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!