How to choose the suitable FNS specification in young patients with femoral neck fracture: A finite element analysis.

Injury

The Second Affiliated Hospital of Guangzhou University of Traditional Chinese Medicine, Guangzhou University of Chinese Medicine, Guangzhou 510006, China.

Published: August 2021

Background: Consensus regarding the optimal approach for the treatment of femoral neck fractures remains lacking. A new internal fixation femoral neck system (FNS) was developed and used in clinical practice. We aimed to investigate the biomechanical outcomes of different types of FNS in the treatment of unstable femoral neck fractures.

Method: In this study, we constructed three different types of unstable femoral neck fractures of Pauwels classification with angles of 50°, 60°, and 70°. We set up four test groups, namely, the one-hole plated FNS group, two-hole plated FNS group, inverted cannulated screw group and triangle cannulated screw group. Under 2100 N axial loads, displacements and the von Mises stress of the femur and internal fixation components were measured for each fracture group.

Results: When the Pauwels angle was 50°or 60°, the one-hole locking plated FNS was as superior as the two-hole plated FNS in terms of femur and internal fixation displacement, and the inverted cannulated screw had slightly better stability than the triangular cannulated screw. However, when the angle increases to 70°, the two-hole locking plate has the minimum displacement, followed by the triangular cannulated screw and inverted cannulated screw, which is the worst displacement for the single-hole locking plate. Regardless of the angle, the two sets of FNS have higher internal fixation stress than the two sets of cannulated screws, which is approximately 1.6-3.0 times that of the cannulated screw group.

Conclusion: From the perspective of biomechanics, we suggest that when the angle of the fracture line is less than 60°, both single-hole locking plated or double-hole locking plated FNS can be used to treat unstable femoral neck fractures. However, when the angle of the fracture line is greater than 70°, we recommend using a double-hole locking plated FNS. This result needs further verification in further clinical studies.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.injury.2021.05.043DOI Listing

Publication Analysis

Top Keywords

cannulated screw
28
femoral neck
24
plated fns
24
internal fixation
16
neck fractures
12
unstable femoral
12
inverted cannulated
12
locking plated
12
fns
10
fns group
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!