Objective: To compare proportions of target range tidal volumes achieved with the self-inflating bag vs. the T-piece in resuscitation of preterm newborns at delivery.

Study Design: This randomized controlled trial was conducted at a tertiary Children's Hospital. 20 preterm infants≤32 weeks' gestational age with no congenital anomalies who needed positive pressure ventilation after birth were enrolled. Positive pressure ventilation was provided with the self-inflating bag or T-piece resuscitator. The primary outcome was proportion of inflations within a target range of 4-8 ml/kg. Chi-square and logistical regression analyses were performed.

Results: In the self-inflating bag (SIB) group 29% of inflations (117/419) and in the T-Piece (TP) group 51% of inflations (300/590) delivered expiratory tidal volume (TVe) of 4-8 ml/kg (p < 65.001). In the SIB group 60% of all inflations (254/419), and in the TP group 35% of all inflations (204/590) delivered TVe < 4 ml/kg (p < 0.001). In the SIB group 11% of all inflations (48/419), and in the TP group, 15% of all inflations (86/590) delivered TVe > 8 ml/kg (p = 0.18). The OR of having expiratory tidal volume of 4-8 ml/kg using the T-piece was 1.8 (CI 1.1-3.1), p = 0.02.

Conclusion: Manual inflations provided by the TP deliver expiratory tidal volumes in the range of 4-8 ml/kg more consistently than SIB.

Download full-text PDF

Source
http://dx.doi.org/10.3233/NPM-210728DOI Listing

Publication Analysis

Top Keywords

self-inflating bag
16
ventilation birth
8
t-piece resuscitator
8
target range
8
bag t-piece
8
positive pressure
8
pressure ventilation
8
assisted ventilation
4
self-inflating
4
birth self-inflating
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!