Background: Few systematic comparative studies of the different methods of physical elastography of the spleen are currently available.

Aim: To compare point shear wave and two-dimensional elastography of the spleen considering the anatomical location (upper, hilar, and lower pole).

Methods: As part of a prospective clinical study, healthy volunteers were examined for splenic elasticity using four different ultrasound devices between May 2015 and April 2017. The devices used for point shear wave elastography were from Siemens (S 3000) and Philips (Epiq 7), and those used for two-dimensional shear wave elastography were from GE (Logiq E9) and Toshiba (Aplio 500). In addition, two different software versions (5.0 and 6.0) were evaluated for the Toshiba ultrasound device (Aplio 500). The study consisted of three arms: A, B, and C.

Results: In study arm A, 200 subjects were evaluated (78 males and 122 females, mean age 27.9 ± 8.1 years). In study arm B, 113 subjects were evaluated (38 men and 75 women, mean age 26.0 ± 6.3 years). In study arm C, 44 subjects were enrolled. A significant correlation of the shear wave velocities at the upper third of the spleen ( = 0.33088, < 0.0001) was demonstrated only for the Philips Epiq 7 device compared to the Siemens Acuson S 3000. In comparisons of the other ultrasound devices (GE, Siemens, Toshiba), no comparable results could be obtained for any anatomical position of the spleen. The influencing factors age, gender, and body mass index did not show a clear correlation with the measured shear wave velocities.

Conclusion: The absolute values of the shear wave elastography measurements of the spleen and the two different elastography methods are not comparable between different manufacturers or models.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8188838PMC
http://dx.doi.org/10.4329/wjr.v13.i5.137DOI Listing

Publication Analysis

Top Keywords

shear wave
28
wave elastography
16
elastography spleen
12
study arm
12
two-dimensional shear
8
point shear
8
ultrasound devices
8
philips epiq
8
aplio 500
8
subjects evaluated
8

Similar Publications

Objective: To investigate whether there is a difference in hardness and hemoglobin concentration changes in the infrapatellar fat pad (IFP) during isometric quadriceps exercise (IQE) in patients with knee osteoarthritis (KOA) between those with and without knee extension limitation.

Design: In this cross-sectional study, data were collected at an orthopedic clinic from March 2022 to April 2023. Among patients diagnosed with KOA, those with knee joint extension range of motion <0° and >0° were defined as the limited group (n ​= ​16) and non-limited group (n ​= ​13), respectively.

View Article and Find Full Text PDF

Biomechanical assessment of ligament maturation after arthroscopic ligament reconstruction of the anterior talofibular ligament.

Orthop Traumatol Surg Res

January 2025

Department of orthopedic surgery, Clinique du Sport, 36 Boulevard Saint-Marcel, 75005 Paris, France.

Background: Many techniques have been described for lateral ankle ligament reconstruction. Although the biomechanical properties of gracilis tendons are different from those of ligaments, the use of a gracilis tendon autograft is a popular option for anatomical reconstruction. Graft maturation and the biomechanical processes over time remain unclear.

View Article and Find Full Text PDF

Exploring the influence of perineal biometrics and stiffness measured by elastography during pregnancy on perineal tears : a pilot study.

J Gynecol Obstet Hum Reprod

January 2025

Nanomedicine Imaging and Therapeutics Laboratory, INSERM EA 4662, University of Franche-Comte, Besançon, France; CHU de Besançon, Service de Gynécologie-Obstétrique, Besançon, France.

Objectives: This study aimed to describe the biometrics and elasticity of the perineal body and the anal sphincter in the ninth month of pregnancy and explore their association with the risk of perineal tears during childbirth.

Methods: In this prospective observational study, pregnant women at 36-40 weeks of gestation were included. Using transperineal 2D-mode ultrasound and shear wave elastography (SWE), we measured the biometrics and stiffness of the perineal body (PB), external anal sphincter (EAS), internal anal sphincter (IAS), and anal mucosa (AM) at rest and during Valsalva maneuvers.

View Article and Find Full Text PDF

Background: The study focuses on the use of multi-parametric ultrasound [gray scale, color Doppler and shear wave elastography (SWE)] to differentiate stable renal allografts from acute graft dysfunction and to assess time-dependent changes in parenchymal stiffness, thereby assessing its use as an efficient monitoring tool for ongoing graft dysfunction. To date, biopsy is the gold standard for evaluation of acute graft dysfunction. However, because it is invasive, it carries certain risks and cannot be used for follow-up monitoring.

View Article and Find Full Text PDF

Myofascial pain is one of the common symptoms in patients with temporomandibular joint disorders (TMD). Occlusal splint (OS) and masticatory muscle trigger point (TP) local injections are primary treatment options. We aimed to investigate the effects of these treatments using clinical and elastography measures.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!