Rationale & Objective: Management of chronic kidney disease mineral and bone disorder requires parathyroid hormone (PTH) concentrations. "Biointact" PTH immunoassays detect "whole" PTH (wPTH), whereas "intact" immunoassays measure PTH plus PTH fragments (iPTH). We aimed to determine whether longitudinal changes in PTH concentrations can be evaluated using biointact and intact immunoassays alike.

Study Design: Open noninterventional longitudinal cohort study.

Setting & Participants: PTH concentrations were measured quarterly up to 5 times in 102 hemodialysis patients.

Predictors & Tests Compared: Age, sex, phosphate levels, and others as clinical predictors for PTH trend. Tests compared were iPTH immunoassays from Siemens and Roche and wPTH immunoassays from Roche and DiaSorin.

Outcomes: PTH concentration trend; regression equations; test bias.

Analytical Approach: Predictive regression-to-the-mean model for PTH slope; Bland-Altman plots, Passing-Bablok regression, and reference change values for test comparisons.

Results: wPTH concentrations were similar with both immunoassays (wPTH-Roche = 11.7 + 0.97 × wPTH-DiaSorin,  = 0.99; mean ± 1.96 SD bias, 8.2 ± 43.3 pg/mL [17.5% ± 40.9%], by Bland-Altman plots). iPTH-Siemens concentrations were higher than iPTH-Roche concentrations (iPTH-Siemens = -5.4 + 1.33 × iPTH-Roche,  = 0.99; mean ± 1.96 SD bias, 84.0 ± 180.2 pg/mL [21.1% ± 29.8%], by Bland-Altman plots). iPTH-Roche and iPTH-Siemens concentrations were 2- and 2.5-fold higher than wPTH concentrations, respectively. Full agreement among all 4 immunoassays in detecting both significant and insignificant changes in PTH concentrations, upward or downward from one quarter to the next, was reached in 87% of consecutive measurements. In a predictive model, baseline PTH concentrations > 199 pg/mL (wPTH-Roche), 204 pg/mL (wPTH-DiaSorin), 386 pg/mL (iPTH-Roche), and 417 pg/mL (iPTH-Siemens) correctly predicted declining PTH concentration trend in 62% to 68% of patients, but age, sex, hemodialysis vintage, and calcium and phosphate levels were no significant predictors.

Limitations: Limited number of immunoassays, only 59 patients attended all quarterly samplings.

Conclusions: wPTH-Roche and wPTH-DiaSorin concentrations were similar, while iPTH was higher than wPTH concentrations. The iPTH-Siemens immunoassay is either higher calibrated or detects more fragments than iPTH-Roche. However, longitudinal PTH concentration changes largely coincided with all tested immunoassays.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8178455PMC
http://dx.doi.org/10.1016/j.xkme.2020.12.015DOI Listing

Publication Analysis

Top Keywords

pth concentrations
16
pth
14
concentrations
12
pth concentration
12
bland-altman plots
12
wpth concentrations
12
immunoassays
9
parathyroid hormone
8
changes pth
8
tests compared
8

Similar Publications

The activated carbon from marigold flowers (MG) was used to make an unlabeled electrochemical immunosensor to determine prostate cancer. MG was synthesized by hydrothermal carbonization and pyrolysis. MG had a large surface area, was highly conductive, and biocompatible.

View Article and Find Full Text PDF

Objective: Vitamin D deficiency (VDD) in children can cause hypocalcaemia and rickets, but the prevalence of these complications and the 25-hydroxyvitamin D (25OHD) concentrations below which they arise is uncertain. We investigated this in children (< 18 years) with 25OHD measurements.

Design, Measurement And Patients: We obtained 25OHD results from the regional laboratory database, alongside albumin-adjusted serum calcium (aCa), parathyroid hormone (PTH) and alkaline phosphatase (ALP) within 6 months of the index 25OHD.

View Article and Find Full Text PDF

Fabrication of bio-mimic nanozyme based on Mxene@AuNPs and molecular imprinted poly(thionine) films for creatinine detection.

Biosens Bioelectron

December 2024

Department of Cogno-Mechatronics Engineering, College of Nanoscience and Nanotechnology, Pusan National University, Busan, 46241, Republic of Korea. Electronic address:

Creatinine (Ctn) is a biomarker for chronic kidney disease (CKD). In this study, a highly sensitive and specific detection method for Ctn based on a molecularly imprinted polymer (MIP) based electrochemical biosensor was developed. Mxene (Mx), which has high absorption properties, was modified using carbon screen-printed electrodes (SPCE).

View Article and Find Full Text PDF

Background: Immunisation in pregnancy against pertussis can reduce severe disease in infancy. There are few data on the safety and immunogenicity of vaccines given to pregnant women living with HIV and their infants. We aimed to describe the safety and immunogenicity of a tetanus-diphtheria-acellular pertussis (TdaP) vaccine containing genetically detoxified pertussis toxin given to pregnant women living with HIV and the effect of the vaccine on infant whole-cell pertussis vaccine responses.

View Article and Find Full Text PDF

Background The complex surgical anatomy and intricate structural arrangement of the thyroid region pose significant challenges for surgeons in identifying the parathyroids and recurrent laryngeal nerve (RLN) during thyroid surgeries. Therefore, it is crucial to develop techniques that enhance the identification of these structures and reduce complications during thyroidectomies. Objective This study intends to assess the efficacy and diagnostic value of Methylene Blue dye and its usefulness in identifying, conserving and minimizing injury to parathyroid glands and recurrent laryngeal nerve during thyroidectomies.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!