Study Design: Screw randomized cadaveric study.

Objective: To compare the accuracy of three-dimensional (3D)-printed drill guides versus additional screw guiding techniques for challenging intra- and extrapedicular screw trajectories.

Summary Of Background Data: Pedicle screw placement can be technically demanding, especially in syndromic scoliosis with limited bone stock. Recently, 3D-printing and virtual planning technology have become available as new tools to improve pedicle screw insertion. Differences in techniques exist, while some focus on guiding the drill, others also actively guide subsequent screws insertion. The accuracy of various 3D-printing-assisted techniques has been studied; however, direct comparative studies have yet to determine whether there is a benefit of additional screw guidance.

Methods: Two cadaveric experiments were conducted to compare drill guides with two techniques that introduce additional screw guiding. The screw guiding consisted of either k-wire cannulated screws or modular guides, which were designed to guide the screw in addition to the drill bit. Screws were inserted intra- or extrapedicular using one of each methods according to a randomization scheme. Postoperative computed tomography scanning was performed and fused with the preoperative planning for detailed 3D screw deviation analysis.

Results: For intrapedicular screw trajectories malpositioning was low (2%) and the modular guides revealed a statistically significant increase of accuracy (P  = 0.05) compared with drill guides. All techniques showed accurate cervical screw insertion without breach. For the extrapedicular screw trajectories both additional screw guiding methods did not significantly (P = 0.09) improve accuracy and malpositioning rates remained high (24%).

Conclusions: In this cadaveric study it was found that the additional screw-guiding techniques are not superior to the regular 3D-printed drill guides for the technically demanding extrapedicular screw technique. For intrapedicular screw insertion, modular guides can improve insertion; however, at cervical levels regular 3D-printed drill guides already demonstrated very high accuracy and therefore there is no benefit from additional screw guiding techniques.

Level Of Evidence: 3.

Download full-text PDF

Source
http://dx.doi.org/10.1097/BRS.0000000000004147DOI Listing

Publication Analysis

Top Keywords

screw guiding
24
drill guides
20
additional screw
20
screw
19
extrapedicular screw
16
screw insertion
16
intra- extrapedicular
12
3d-printed drill
12
modular guides
12
guiding
8

Similar Publications

Background: Coronary sinus (CS) lead placement in persistent left superior vena cava (PLSVC) cases is challenging because of the poor backup force of the guiding catheter within the enlarged CS. Active fixation Quadripolar leads (Attain Stability™ Quad 4798, Medtronic) can expand choice to CS branches with limited access; however, no cases of anchoring to the main body of the CS have been published to date.

Case Summary: We describe a case of cardiac resynchronization therapy pacemaker upgrade in a 79-year-old female who developed pacing-induced cardiomyopathy after pacemaker implantation via the right superior vena cava (SVC) for atrioventricular block eight years ago wherein PLSVC was revealed during the procedure.

View Article and Find Full Text PDF

[Treatment with TOPS for short femoral stump].

Oper Orthop Traumatol

January 2025

Klinik für Unfall‑, Hand und Wiederherstellungschirurgie, Universitätsmedizin Rostock, Schillingallee 35, 18057, Rostock, Deutschland.

Objective: Treatment with transcutaneous osseointegrated prosthesis systems (TOPS) for short femoral amputation stumps aims to restore independent walking ability after proximal femoral amputation by direct bone-guided prosthesis anchorage. This cannot be safely achieved with conventional socket prostheses due to the mechanically inadequate socket contact surface.

Indications: Treatment of patients with short transfemoral stumps who cannot be mobilized sufficiently with conventional socket prostheses.

View Article and Find Full Text PDF

With the emergence of numerous classifications, surgical treatment for adolescent idiopathic scoliosis (AIS) can be guided more effectively. However, surgical decision-making and optimal strategies still lack standardization and personalized customization. Our study aims to devise proper deep learning (DL) models that incorporate key factors influencing surgical outcomes on the coronal plane in AIS patients to facilitate surgical decision-making and predict surgical results for AIS patients.

View Article and Find Full Text PDF

Background: This study aims to demonstrate the 3-dimensional pattern of the nutrient vessels of the triquetrum using micro-computed tomography and determine the safe region for screw applications to the bone.

Methods: Seven fresh frozen cadavers' ulnar and radial arteries were injected with Microfil MV-117. Triquetral bones were dissected and examined using micro-computed tomography.

View Article and Find Full Text PDF

Background: Robotic assistance has become increasingly prevalent in spinal surgery in recent years, emerging as a tool to increase accuracy and precision and lower complication rates and radiation exposure. The 7 and 8 Annual Seattle Science Foundation (SSF) Robotics Courses showcased presentations and demonstrations from some of the field's most experiences leaders on latest topics in robotics and spinal surgery, including cutting-edge preoperative planning technologies, augmented reality (AR) in the operating room, cervical fusion with transpedicular screws, and neuro-oncologic management. We provide a scoping review of the use of robotics technology in spinal surgery featuring highlights from the 7 and 8 Annual SSF Robotics Courses.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!