Validation of an assessment tool for estimation of abdominal aortic aneurysm compression in diagnostic ultrasound.

Ultrasonics

Copenhagen Academy for Medical Education and Simulation (CAMES), Teilum, Blegdamsvej 9, 2100 Copenhagen, Denmark; Department of Vascular Surgery, Rigshospitalet, Blegdamsvej 9, 2100 Copenhagen, Denmark; Department of Clinical Medicine, Faculty of Health and Medical Sciences, University of Copenhagen, Blegdamsvej 3B, 2200 Copenhagen, Denmark.

Published: September 2021

The study investigated ultrasound (US) transducer push, tantamount to applied transducer pressure, during abdominal aortic aneurysm (AAA) US scanning in a simulated non-clinical setup. During an assessment of maximal AAA diameter on a three-dimensional print-based AAA phantom, US transducer push varied as much as 2000% (range: 0.52-12.45 kPa) amongst 16 experienced sonographers. The mean transducer push was 5.54 ± 3.91 kPa (CV = 0.71). Deformation of a standardized gel-pad allowed for transducer push calculation based on US images; Young's modulus of the gel-pad was estimated to 44,26 N/m. The method is theoretically validated in a safe and non-clinical environment. Future investigations with the aim of clinical validation of the gel-pad principle on AAA patients are suggested, including the objectification of the magnitude of an eventual transducer push-related error during US AAA diameter measurement.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ultras.2021.106484DOI Listing

Publication Analysis

Top Keywords

transducer push
16
abdominal aortic
8
aortic aneurysm
8
aaa diameter
8
transducer
6
aaa
5
validation assessment
4
assessment tool
4
tool estimation
4
estimation abdominal
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!