Aims: The present study aims at analyzing the role of a preimplantation 12-lead electrocardiogram (ECG) on the prediction of inappropriate S-ICD® episodes.
Methods: N=116 screened patients (pts) with an S-ICD® and a follow-up of at least 6 months were included. A preimplantation 12-lead ECG (50 mm/s, 10 mm/mV) was analyzed with regard to QRS and T-wave amplitude, T wave concordance or discordance and QRS/T wave ratio in all 12 leads. To ensure an exact determination of parameters Datinf® Measure software was used. Results were correlated to the occurrence of oversensing of cardiac signals during follow-up.
Results: N = 116 pts. (63,8% male, mean age 40,9 ± 15,5 years) were included (primary prevention in 47.4% of pts). The most frequent cardiac diseases were hypertrophic cardiomyopathy (HCM) in n = 25 (21,6%), electrical heart disease in n = 20 (17,2%), and dilated cardiomyopathy in n = 17 (14,7%). Mean follow-up was 740 ± 549 days. During follow- up n = 17 (14.7%) pts. experienced n = 27 inappropriate episodes due to T-wave oversensing. Besides HCM (OR 6.16, CI 1.79-21.15, p = 0.004) a discordance of QRS to T-wave in lead I (OR 6.5, CI 1.86-22.67, p = 0.003) was found to be a strong predictor for inappropriate shocks. In multivariate analysis the pts. with a combination of both had an 8.4-fold higher risk of misclassification of intracardiac signals (p = 0.003) with consecutive inappropriate therapy.
Conclusion: A discordance of QRS to T-wave in lead I turned out to be a strong predictor for future inappropriate shocks in a typical S-ICD® cohort with special impact on HCM pts.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jelectrocard.2021.05.011 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!