Background: Nickel-titanium (NiTi) open coils are common tools in fixed orthodontic treatments to apply light and continuous forces. Their favorable mechanical properties and their relatively high cost have prompted many clinicians to decide to reuse them.

Aim: The present study aimed to evaluate the effect of 10% hydrochloric acid (HCl) and 5.25% sodium hypochlorite (NaClO) solution and sterilization, on the unloading force of NiTi open-coils.

Materials And Methods: This experimental study consisted of 160 NiTi open coils from two brands of Highland Metals and Masel were provided. Each group was randomly divided into four subgroups: control, three rounds of immersion, three rounds of autoclave sterilization, and three rounds of immersion and autoclaving. Each round of immersion process included 1 min in 10% HCl, followed by 5 min in 5.25% NaClO. Spring was compressed for 4 mm. Then, the unloading forces of each spring were recorded at 4, 3, 2, and 1 mm of compression, respectively. The data were analyzed with SPSS ( < 0.05), using one-way ANOVA and independent -test.

Results: One-way ANOVA showed a significant difference in the 2-mm compression of Group 1 and 3-mm compressions of Groups 1 and 2 compared to the control group in the Highland metals brand ( < 0.04, < 0.014, and < 0.007, respectively). There was no significant difference between the different compressed lengths and groups in the Masel brand.

Conclusion: One-minute immersion in 10% HCL and 5-min immersion in 5.25% NaOCl, followed by autoclave sterilization, even after repeating the process three times, did not significantly change the force properties of NiTi open springs. Therefore, the re-use of these springs with the above-mentioned method as preparation can be recommended.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8122694PMC

Publication Analysis

Top Keywords

autoclave sterilization
12
open coils
12
niti open
12
three rounds
12
hydrochloric acid
8
sodium hypochlorite
8
highland metals
8
rounds immersion
8
10% hcl
8
one-way anova
8

Similar Publications

<b>Background and Objective:</b> Todolo coffee (<i>Coffea arabica</i> L. var. typica) is the oldest commercially grown coffee in the Toraja region of South Sulawesi and is currently at risk of extinction.

View Article and Find Full Text PDF

Sterilization is required for any biomedical device intended to be used in contact with the human body. Several studies have reported alterations in the bulk and surface properties of such devices after repeated sterilization cycles. These surface modifications may influence other clinical parameters.

View Article and Find Full Text PDF

A Comparative In-Vitro Study of the Effectiveness of Several Methods of Sterilizing Endodontic Files.

Cureus

November 2024

Faculty of health and medical technology, Department of Medical Laboratory Technology, Al-Ayen University, Nasiriyah, IRQ.

Background Successful endodontic treatment depends on several elements. The most important of which is the elimination of microorganisms within the root canal system. To achieve this, sterility must be maintained during all steps of the treatment.

View Article and Find Full Text PDF

Disinfection practices in dental settings are fundamental to clinical safety, playing a pivotal role in preventing cross-infections and protecting the health of patients and healthcare professionals. This article examines the key components of effective disinfection, based on evidence-based protocols developed by international organizations such as the WHO and the U.S.

View Article and Find Full Text PDF

Background: Enterococcus faecalis (E. faecalis) is the most common microorganism responsible for post-endodontic infections, which can penetrate deep into dentinal tubules. This study aimed to compare the survival of E.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!