Stent deployment in a calcified coronary artery is often associated with suboptimal outcomes such as stent underexpansion and malapposition. Post-dilation after stent deployment is commonly used for optimal stent implantation. There is no guideline for choosing the post-dilation balloon diameter and inflation pressure. In this work, ex-vivo/in-silico experiments were performed to investigate the efficacy of post-dilation balloon diameter and inflation pressure in improving the stent expansion in a calcified lesion. Post-dilations with three balloon diameters (3 mm, 3.5 mm, and 4 mm) were performed. For each balloon diameter, three inflation pressures (10 atm, 20 atm, and 30 atm) were sequentially applied. In ex-vivo experiments, optical coherence tomography images were acquired during the stenting procedure, i.e., pre- and post-deployment of 3 mm diameter stent, as well as after each post-dilation. The results from in-silico experiments were compared with ex-vivo experiments in terms of lumen area. In addition, stretch ratio analysis was developed to predict the stent-induced lumen area, along with the strain analysis and the in-silico experiments. Results have shown that target lumen area could be achieved with an oversized nominal balloon diameter of +0.5 mm (i.e., 0.5 mm greater than reference lumen diameter) at an inflation pressure of 20 atm. After each post-dilation, fibrotic tissue demonstrated a larger strain, contributing to improved lumen gain. However, minimal changes were observed in calcification. Moreover, a strong correlation (R = 0.95) between the stretch ratio of fibrotic tissue and lumen area after each post-dilation was observed. This indicated that the morphology of the fibrotic tissue could be a potential marker to predict the lumen gain. The detailed mechanistic quantifications of a single lesion cannot be generalized to all clinical cases. However, this work could be used to provide a fundamental understanding of the post-dilations, to develop experimental protocols for producing generalized guidelines, and to exploit their potential for optimal pre- and post-stent strategies.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jmbbm.2021.104609DOI Listing

Publication Analysis

Top Keywords

balloon diameter
16
lumen area
16
diameter inflation
12
inflation pressure
12
fibrotic tissue
12
improving stent
8
stent expansion
8
expansion calcified
8
calcified coronary
8
coronary artery
8

Similar Publications

Background: Drug-coated balloons (DCB) can decrease the incidence of restenosis in the treatment of intracranial atherosclerotic stenosis (ICAS). This study aimed to assess the safety and efficacy of submaximal angioplasty with DCB dilation compared with aggressive angioplasty in patients with symptomatic ICAS.

Methods: This study prospectively and consecutively enrolled patients with symptomatic ICAS who underwent DCB angioplasty between January 2021 and December 2023.

View Article and Find Full Text PDF

Background: Endoscopic treatment for second primary malignancies after esophagectomy has been increasingly performed; however, evidence regarding the outcomes of endoscopic submucosal dissection (ESD) for superficial cancer of the remnant esophagus after esophagectomy (SCREE) is limited.

Methods: We retrospectively extracted cases of ESD for SCREE from our institutional database, which included 739 consecutive esophageal ESD procedures performed between January 2009 and September 2023. Information on prior treatment, clinical features of the lesions, and outcomes was evaluated.

View Article and Find Full Text PDF

Pancreaticojejunostomy (PJ) is a critical step in pancreaticoduodenectomy (PD), often complicated by the risk of postoperative pancreatic fistula (POPF). This video report demonstrates a novel robotic PJ technique employing a self-expandable metallic stent. The method involves the use of the Da Vinci Xi robotic system and the WallFlex™ Biliary RX Stent for improved anastomotic support, particularly in high-risk cases defined by soft pancreatic texture and narrow duct diameter (<3 mm).

View Article and Find Full Text PDF

Objective: The gold standard imaging modality for assessing crural arterial disease is digital subtraction angiography (DSA). Using DSA, the operator can estimate the diameter and degree of stenosis and determine the diameter of the balloon. Since these measures are estimates, it allows for interobserver and intraobserver variability.

View Article and Find Full Text PDF

Novel Real-Time Fluoroscopic Assessment Method of Transcatheter Heart Valve Expansion Following Balloon-Expandable TAVR.

Circ Cardiovasc Interv

January 2025

Centre for Cardiovascular Innovation, University of British Columbia, Vancouver, Canada. (A.H., J.J., S.O., K.M., J.A.L., P.B., D.A.W., S.L.S., J.G.W., J.S.).

Background: Transcatheter heart valve (THV) underexpansion after transcatheter aortic valve replacement may be associated with worse outcomes. THV expansion can be assessed fluoroscopically using a pigtail for calibration; however, the accuracy of this technique specific to transcatheter aortic valve replacement is unknown. We assessed the accuracy and reproducibility of a novel fluoroscopic method to assess THV expansion using the THV commissural post for calibration.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!