Significance: Fixation disparity tests with various characteristics of fusional stimulus are very important for assessment of decompensated heterophoria. The results suggest that there was no reasonable agreement between the fixation disparity curve's parameters of the modified near Mallett unit and the Sheedy disparometer.

Purpose: The purpose of this study was to determine the agreement of the fixation disparity curve parameters between the modified near Mallett unit and the Sheedy disparometer in patients with decompensated near heterophoria.

Methods: A total of 147 young adults (mean age, 22.7 ± 4.8 years) participated in this cross-sectional study. After applying the exclusion criteria, the statistical analysis was done on the data of 134 individuals. All participants underwent preliminary optometric examinations including the measurement of uncorrected and best-corrected visual acuity, objective and subjective refraction, and unilateral and alternating cover tests. The fixation disparity was evaluated using the modified near Mallett unit and the Sheedy disparometer at 40 cm, and forced-vergence fixation disparity curves were generated.

Results: There were statistically significant differences in the fixation disparity curve parameters (except the center of symmetry) between the two devices. The median fixation disparity measured by the Sheedy disparometer was more positive compared with the modified near Mallett unit (toward more esodisparity or less exodisparity). The median associated phoria measured by the Sheedy disparometer was more positive compared with the Mallett unit. Also, the slope of the curve obtained by the Mallett unit was steeper. The wide limits of agreement indicated the poor agreement of all fixation disparity curve parameters between the two instruments.

Download full-text PDF

Source
http://dx.doi.org/10.1097/OPX.0000000000001708DOI Listing

Publication Analysis

Top Keywords

fixation disparity
36
mallett unit
24
agreement fixation
16
disparity curve
16
modified mallett
16
sheedy disparometer
16
unit sheedy
12
curve parameters
12
disparity
9
fixation
8

Similar Publications

To investigate the pattern and threshold of physiological growth, defining as axial length (AL) elongation that results in little refraction progression, among Chinese children and teenagers, a total of 916 children aged between 7 and 18 years from a 6-year longitudinal cohort study were included for analysis. Ocular biometry, cycloplegic refraction and demographic data were obtained annually. Physiological growth was calculated based on myopic progression and Gullstrand eye model, respectively.

View Article and Find Full Text PDF

Objective: To compare the short-term clinical efficacy and safety of closed reduction with Kirschner wire fixation versus open reduction with plate fixation for treating osteoporotic Colles' fractures in middle-aged and elderly patients.

Methods: Between January 2018 and January 2023, 119 patients with Colles fractures were retrospectively analyzed, including 39 males and 80 females, aged from 48 to 74 years old with an average of(60.58±6.

View Article and Find Full Text PDF

Barriers to Healthcare Access and Utilization Among Racial and Ethnic Minorities with Noninfectious Uveitis.

Ophthalmology

January 2025

Department of Ophthalmology & Vision Science, Tschannen Eye Institute, University of California, Davis, Sacramento, California, United States. Electronic address:

Objective: To investigate barriers to healthcare access and utilization among patients with noninfectious uveitis (NIU) across different racial and ethnic groups.

Design: Retrospective cross-sectional study.

Subjects: Participants diagnosed with chronic NIU in the nationwide All of Us Research Program.

View Article and Find Full Text PDF

Anterior cruciate ligament reconstruction aims to improve knee stability and range of motion. The AperFix system consists of polymer components, and fixed-loop fixation is an established endoscopic technique. Our aim in this study was to compare the long-term clinical and radiological results of AperFix and fixed-loop fixation and to prove that the long-term results of the AperFix fixation method are at least as good as those of the fixed loop device.

View Article and Find Full Text PDF

Binocular vision requires that the brain integrate information coming from each eye. These images are combined (fused) to generate a meaningful composite image. Differences between images, within a range, provide useful information about depth (stereopsis).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!