Purpose: The purpose of this work is to (a) demonstrate the feasibility of delivering a spread-out Bragg peak (SOBP) proton beam in ultra-high dose rate (FLASH) using a proton therapy synchrocyclotron as a major step toward realizing an experimental platform for preclinical studies, and (b) evaluate the response of four models of ionization chambers in such a radiation field.

Methods: A clinical Mevion HYPERSCAN synchrocyclotron was adjusted for ultra-high dose rate proton delivery. Protons with nominal energy of 230 MeV were delivered in pulses with temporal width ranging from 12.5 μs to 24 μs spanning from conventional to FLASH dose rates. A boron carbide absorber and a range modulator block were placed in the beam path for range modulation and creating an SOBP dose profile. The radiation field was defined by a brass aperture with 11 mm diameter. Two Faraday cups were used to determine the number of protons per pulse at various dose rates. The dosimetric response of two cylindrical (IBA CC04 and CC13) and two plane-parallel (IBA PPC05 and PTW Advanced Markus ) ionization chambers were evaluated. The dose rate was measured using the plane-parallel ionization chambers. The integral depth dose (IDD) was measured with a PTW Bragg Peak ionization chamber. The lateral beam profile was measured with EBT-XD radiochromic film. Monte Carlo simulation was performed in TOPAS as the secondary check for the measurements and as a tool for further optimization of the range modulators' design.

Results: Faraday cups measurement showed that the maximum protons per pulse is 39.9 pC at 24 μs pulse width. A good agreement between the measured and simulated IDD and lateral beam profiles was observed. The cylindrical ionization chambers showed very high ion recombination and deemed not suitable for absolute dosimetry at ultra-high dose rates. The average dose rate measured using the PPC05 ionization chamber was 163 Gy/s at the pristine Bragg peak and 126 Gy/s at 1 cm depth for the SOBP beam. The SOBP beam range and modulation were measured 24.4 mm and 19 mm, respectively. The pristine Bragg peak beam had 25.6 mm range. Simulation results showed that the IDD and profile flatness can be improved by the cavity diameter of the range modulator and the number of scanned spots, respectively.

Conclusions: Feasibility of delivering protons in an SOBP pattern with >100 Gy/s average dose rate using a clinical synchrocyclotron was demonstrated. The dose heterogeneity can be improved through optimization of the range modulator and number of delivered spots. Plane-parallel chambers with smaller gap between electrodes are more suitable for FLASH dosimetry compared to the other ion chambers used in this work.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.15021DOI Listing

Publication Analysis

Top Keywords

bragg peak
20
dose rate
20
ionization chambers
16
ultra-high dose
12
dose rates
12
range modulator
12
dose
11
spread-out bragg
8
clinical synchrocyclotron
8
feasibility delivering
8

Similar Publications

Purpose: Proton FLASH has been investigated using cyclotron and synchrocyclotron beamlines but not synchrotron beamlines. We evaluated the impact of dose rate (ultra-high [UHDR] vs. conventional [CONV]) and beam configuration (shoot-through [ST] vs.

View Article and Find Full Text PDF

Variable relative biological effectiveness (RBE) of carbon radiotherapy may be calculated using several models, including the microdosimetric kinetic model (MKM), stochastic MKM (SMKM), repair-misrepair-fixation (RMF) model, and local effect model I (LEM), which have not been thoroughly compared. In this work, we compared how these four models handle carbon beam fragmentation, providing insight into where model differences arise. Monoenergetic and spread-out Bragg peak carbon beams incident on a water phantom were simulated using Monte Carlo.

View Article and Find Full Text PDF

Fiber Bragg gratings (FBGs) inscribed by UV light and different femtosecond laser techniques (phase mask, point-by-point, and plane-by-plane) were exposed-in several irradiation cycles-to accumulated high doses of gamma rays (up to 124 MGy) and neutron fluence (8.7 × 10/cm) in a research-grade nuclear reactor. The FBG peak wavelengths were measured continuously in order to monitor radiation-induced shifts.

View Article and Find Full Text PDF

Background: In proton radiotherapy, the steep dose deposition profile near the end of the proton's track, the Bragg peak, ensures a more conformed deposition of dose to the tumor region when compared with conventional radiotherapy while reducing the probability of normal tissue complications. However, uncertainties, as in the proton range, patient geometry, and positioning pose challenges to the precise and secure delivery of the treatment plan (TP). In vivo range determination and dose distribution are pivotal for mitigation of uncertainties, opening the possibility to reduce uncertainty margins and for adaptation of the TP.

View Article and Find Full Text PDF

Pancreatic ductal adenocarcinoma (PDAC) is notably resistant to conventional chemotherapy and radiation treatment. However, clinical trials indicate that carbon ion radiotherapy (CIRT) with concurrent gemcitabine is effective for unresectable locally advanced PDAC. This study aimed to identify patient characteristics predictive of CIRT response.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!