Increasing the height of the anterior lip on a tibial insert in a posterior stabilized knee prosthesis has little effect on the wear rate.

Med Eng Phys

School of Biological Science and Medical Engineering, Beihang University, Beijing 100083, China; School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, 200030, China; Beijing Advanced Innovation Center for Biomedical Engineering, Beihang University, Beijing 100083, China. Electronic address:

Published: May 2021

A high anterior lip on a total knee prosthesis is an effective way of reducing anterior translation, but the effect on joint wear is unclear. Using finite element analysis (FEA), this study quantitatively compared wear rates and anterior contact stresses in three posterior stabilized knee prostheses with different heights for the anterior lip during six daily activities (walking, stair ascent, stair descent, sit-to-stand, pivot turn and crossover turn). The wear rate and location of maximum wear depth were similar for the three lip heights tested, but the knee with the highest anterior lip also showed slight anterior wear scaring due to articular contact stress during swing phase, which was highly dependent on the shape of the contact interface. This study illustrates that tibial inserts with a high anterior lip maintain a wear rate similar to moderate and low lip posterior stabilized designs.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.medengphy.2021.03.007DOI Listing

Publication Analysis

Top Keywords

anterior lip
20
posterior stabilized
12
wear rate
12
anterior
8
stabilized knee
8
knee prosthesis
8
high anterior
8
lip
7
wear
7
increasing height
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!