Article Abstract

Purpose: To compare the induced corneal stromal bed roughness measured with atomic force microscopy (AFM) after LASIK flap creation with the IntraLase 60 kHz and the VisuMax femtosecond laser platforms.

Methods: Three freshly enucleated porcine eyes were operated with each femtosecond laser in this experimental study. Standard LASIK treatment parameters were used for the experiment. After LASIK flap creation, the corneal stromal roughness was assessed using a JPK NanoWizard II® AFM in contact mode immersed in liquid. Olympus OMCL-RC800PSA commercial silicon nitride cantilever tips were used. Surface measurements were made in 10 regions of the central cornea of each sample measuring 20 x 20 microns, at 512 x 512 point resolution. Roughness was measured using the root-mean-square (RMS) value within the given regions.

Results: Measurements from 30 regions of the 3 eyes (10 measurements per eye) in the Intralase (FS1) group, and 30 regions of the 3 eyes (10 measurements per eye) in the VisuMax (FS2) group were analyzed. There was a statistically significant difference in mean ± standard deviation RMS values between the FS1 and the FS2 groups (360 ± 120 versus 230 ± 100 nm respectively; P< 0.00001).

Conclusion: This AFM study indicates that the surface of the stromal bed after LASIK flap creation is smoother in the FS2 group than the FS1 group.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8158881PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0252449PLOS

Publication Analysis

Top Keywords

corneal stromal
12
femtosecond laser
12
lasik flap
12
flap creation
12
stromal roughness
8
atomic force
8
force microscopy
8
stromal bed
8
roughness measured
8
measurements regions
8

Similar Publications

Purpose: To describe a rare case of infectious keratitis secondary to Brevundimonas diminuta, a gram-negative bacillus with fluoroquinolone resistance and rare clinical isolation.

Methods: A 50-year-old man with contact lens overuse presented with a large corneal ulcer and hand motion visual acuity. Initial treatment with fortified topical tobramycin and vancomycin yielded slow improvement, and initial culture grew Staphylococcus epidermidis, Staphylococcus hominis, and Corynebacterium bovis.

View Article and Find Full Text PDF

PCR-based diagnosis and clinical insights into parasitic keratitis.

J Microbiol Immunol Infect

January 2025

Department of Ophthalmology, National Cheng Kung University Hospital, College of Medicine, National Cheng Kung University, Tainan, Taiwan. Electronic address:

Purpose: This retrospective study aimed to investigate demographic characteristics, predisposing factors, and clinical outcomes in patients with parasitic keratitis.

Methods: Medical records of patients with molecularly confirmed Acanthamoeba or microsporidia, identified through corneal scraping specimens (collected between September 21, 2017, and June 27, 2023), were reviewed. Demographic data, clinical profiles, such as symptom duration before confirmed diagnosis, antiviral treatment pre-diagnosis, contact lens use, tap water and soil contamination, ocular trauma, and treatment regimens, were analyzed.

View Article and Find Full Text PDF

Distinct Ocular Surface Microbiome in Keratoconus Patients Correlate With Local Immune Dysregulation.

Invest Ophthalmol Vis Sci

January 2025

GROW Research Laboratory, Narayana Netralaya Foundation, Bangalore, India.

Purpose: Keratoconus (KC) is characterized by irregular astigmatism along with corneal stromal weakness and is associated with altered immune status. Tissue resident microbiomes are known to influence the immune status in other organs, but such a nexus has not been described in ocular conditions. Therefore, we examined the ocular surface microbiome of patients with KC and correlated it to the immune cell and tear molecular factor profiles.

View Article and Find Full Text PDF

Background: Glaucoma, particularly open-angle glaucoma (OAG), is a leading cause of irreversible blindness, associated with optic nerve damage, retinal ganglion cell death, and visual field defects. Corneal biomechanical properties and cellular components, such as corneal nerve and keratocyte densities assessed by in vivo confocal microscopy (IVCM), may serve as biomarkers for glaucoma progression. This study aimed to explore the relationship between corneal nerve parameters, keratocyte density, and optical coherence tomography (OCT)-derived retinal nerve fiber layer (RNFL) thickness in primary open-angle glaucoma (POAG) patients and controls.

View Article and Find Full Text PDF

The study aimed to compare the effects of different types of excimer laser keratectomy on rabbit corneas and to identify the optimal disease model for corneal ectasia. Additionally, investigating the structural and molecular alterations in the novel disease model helped explore the mechanisms underlying biomechanical cues in corneal ectasia. 2.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!