Patient-to-robot registration: The fate of robot-assisted stereotaxy.

Int J Med Robot

Department of Neurosurgery and Neurotechnology, Neurosurgical Clinic, Eberhard Karls University, Tuebingen, Germany.

Published: October 2021

AI Article Synopsis

Article Abstract

Background: Robot-assisted stereotaxy (RAS) promises higher stereotactic accuracy (SA) and time efficiency (TE) than frame-based stereotaxy. However, both aspects are attributed to the problem of patient-to-robot registration.

Objective: To examine different registration techniques regarding their SA and TE.

Methods: This study enrolled 57 patients undergoing RAS with bone fiducial registration (BFR) or laser surface registration (LSR). SA was measured by the entry point error (EPE). Additionally, predictors of SA (registration error [RegE], distance-to-registration plane [DTC]) and TE (imaging, skin-to-skin) were assessed.

Results: The mean SA was 1.0 ± 0.8 mm. BFR increased SA by reducing RegE and DTC. In LSR, EPE depended on DTC (face and forehead) with highest accuracy for DTC ≤100 mm. CT-based LSR exerted a higher SA than MR-based LSR. In BFR, TE was confined by the additional imaging.

Conclusion: Every registration technique counteracts one of the promises of RAS. New solutions are needed to increase the acceptance of RAS in neurosurgery.

Download full-text PDF

Source
http://dx.doi.org/10.1002/rcs.2288DOI Listing

Publication Analysis

Top Keywords

robot-assisted stereotaxy
8
registration
5
patient-to-robot registration
4
registration fate
4
fate robot-assisted
4
stereotaxy background
4
background robot-assisted
4
ras
4
stereotaxy ras
4
ras promises
4

Similar Publications

Robot-Assisted 2D Fluoroscopic Needle Placement-A Phantom Study.

Diagnostics (Basel)

August 2024

Interventional Oncology-Stereotaxy & Robotics (SIP), Department of Radiology, Medical University Innsbruck, Anichstr. 35, 6020 Innsbruck, Austria.

Rationale And Objectives: To evaluate the targeting accuracy of a novel robot-assisted guidance technique relying on one pair of 2D C-arm images.

Material And Methods: In total, 160 punctures were carried out semi-automatically by using a novel robotic device. The needle's paths were planned based on one pair of 2D fluoroscopic images from different angles.

View Article and Find Full Text PDF
Article Synopsis
  • Interventional oncology involves procedures like tumor characterization and destruction, which can be complex due to difficult access and risk factors.
  • Various robotic systems have been developed to minimize human error and variability during these procedures, aiming to improve accuracy and reduce risks associated with radiation exposure.
  • This review compiles evidence on the clinical effectiveness and safety of these robotic tools in treating nonhepatic malignancies, highlighting their potential benefits for patient outcomes.
View Article and Find Full Text PDF

Stereoelectroencephalography (SEEG) is the gold standard to investigate the epileptic network in cases of drug-resistant epilepsy. Robot-assisted SEEG is increasingly being used; however, its installation process in the operating room is more difficult than that of the stereotactic frame procedure. New robotic tools and 3D intraoperative imaging ease the setup while achieving the same mechanical precision and a lower complication rate.

View Article and Find Full Text PDF

Background And Objectives: Despite frequent use, stereotactic head frames require manual coordinate calculations and manual frame settings that are associated with human error. This study examines freestanding robot-assisted navigation (RAN) as a means to reduce the drawbacks of traditional cranial stereotaxy and improve targeting accuracy.

Methods: Seven cadaveric human torsos with heads were tested with 8 anatomic coordinates selected for lead placement mirrored in each hemisphere.

View Article and Find Full Text PDF

Introduction: Diffuse midline glioma (DMG) of the pons occurs in pediatric patients and carries a dismal prognosis. Biopsy is not necessary for diagnosis but provides information, particularly H3K27M status, with prognostic implications. Additionally, biopsy information may open therapeutic options such as clinical trials that require mutation status.

View Article and Find Full Text PDF