Walrus Balloon Guide Catheter for Stroke Intervention: Technical Considerations and Clinical Outcomes.

World Neurosurg

Department of Neurosurgery, Jacobs School of Medicine and Biomedical Sciences, State University of New York at Buffalo, Buffalo, New York, USA; Department of Bioinformatics, Jacobs School of Medicine and Biomedical Sciences, State University of New York at Buffalo, Buffalo, New York, USA; Department of Neurosurgery, Gates Vascular Institute at Kaleida Health, Buffalo, New York, USA; University of California Santa Barbara, Santa Barbara, California, USA; Jacobs Institute, Buffalo, New York, USA. Electronic address:

Published: August 2021

Background: The use of balloon guide catheters (BGCs) for proximal flow arrest during neurointerventional procedures has been limited owing to the incompatibility of BGCs with large-bore aspiration catheters and difficulty in device navigation. The objective of our study was to describe the use of the Walrus catheter (Q'Apel Medical, Fremont, California, USA), a new 8F BGC, with a variety of aspiration catheters and procedures requiring flow arrest.

Methods: Consecutive cases using Walrus BGCs for proximal flow arrest during mechanical thrombectomy for acute stroke cases were recorded. The procedure indication, vessel occlusion site, technique, first-pass effect (modified thrombolysis in cerebral infarction score of 2C or 3 after the first recanalization attempt), and complications were recorded and evaluated statistically.

Results: Our study included 57 patients who had undergone mechanical thrombectomy. In addition to mechanical thrombectomy, the Walrus BGC was used in conjunction with the following techniques: stent retrieval in 2 patients (3.5%), stent retrieval followed by aspiration (Solumbra technique) in 41 (71.9%), and aspiration followed by stent retrieval in 14 patients (24.6%). Eight different aspiration catheters were used in 56 of these 57 procedures. The first-pass effect (modified thrombolysis in cerebral infarction score of 2C and 3) was achieved in 36 (63.2%) of 57 procedures. Two patients (3.5%) had experienced intraoperative complications (symptomatic intracerebral hemorrhage) and one patient (1.8%) had died in-hospital.

Conclusions: Our results have demonstrated that the Walrus BGC is a highly navigable 8F guide catheter compatible with most available aspiration catheters. Owing to its compatibility with most available aspiration catheters and ease of use, the Walrus BGC is a valuable addition to the tools available for mechanical thrombectomy.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.wneu.2021.05.051DOI Listing

Publication Analysis

Top Keywords

aspiration catheters
20
mechanical thrombectomy
16
walrus bgc
12
stent retrieval
12
balloon guide
8
guide catheter
8
bgcs proximal
8
proximal flow
8
flow arrest
8
catheters procedures
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!