A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Does the Anchor Eye Alignment Affect the Performance of the Rotator Cuff Repair? A Biomechanical Study. | LitMetric

Does the Anchor Eye Alignment Affect the Performance of the Rotator Cuff Repair? A Biomechanical Study.

Arthrosc Sports Med Rehabil

Laboratory of Biomechanical Engineering, University Hospital, Department of Mechanical Engineering, Federal University of Santa Catarina, Florianópolis, Brazil.

Published: April 2021

Purpose: To investigate the metal screw-in anchor failure mode and load to failure for 2 different eyelet alignments after anchor insertion in ovine humeri.

Methods: Sixteen ovine humeri were dissected, and a 5-mm metal anchor with 2 nonabsorbable polyblend polyethylene sutures was inserted into them in the proximal position of the greater tuberosity. The alignment of the anchors after insertion was adjusted to make 2 test groups, each with 8 specimens: In group 1, the anchor eyelets were malpositioned, whereas in group 2, the anchor eyelets were aligned according to the manufacturer's instructions. After insertion, cyclic tests from 10 N to 180 N were performed with a frequency of 1 Hz for 200 cycles; specimens were then loaded to failure to evaluate the maximum load of the system and observe the associated failure mode.

Results: The mean ultimate failure load in group 2 was not significantly different from that in group 1 ( = .472).

Conclusions: For metallic screw-in suture anchors, the alignment of the eyelet does not change the failure mode and the load to failure after cyclic loading of the bone-anchor-suture system in ovine humeri.

Clinical Relevance: Our results indicate that on the basis of this anchor model, the position of the eyelet in the greater tuberosity does not interfere directly with the biomechanical performance of the system.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8128988PMC
http://dx.doi.org/10.1016/j.asmr.2020.10.005DOI Listing

Publication Analysis

Top Keywords

failure mode
8
mode load
8
load failure
8
greater tuberosity
8
group anchor
8
anchor eyelets
8
anchor
7
failure
7
anchor eye
4
eye alignment
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!